Meditherm Inc.
5353 Wyoming Blvd Ne Ste D, Albuquerque, New Mexico, United States
FEI 3006111110
FDA has 2 inspections on file for Meditherm Inc. in Albuquerque, New Mexico (FEI 3006111110), from 2018 to 2022. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
The most recent OAI inspection ended on May 30, 2018. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.20(c) and 21 CFR 820.22.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 2, 2018–2022
- 1 OAI · 0 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest FY2023
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been established.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of procedures, or not maintained
Document control procedures have not been established.
Records of complaint investigation
Records of complaint investigations do not include required information.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Retention period
Required records are not retained for the design and expected life of the device.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Show 10 more citations
Design history file
The design history file was not established and does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Design control - no procedures
Procedures for design control have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
DMR - not or inadequately maintained
A device master record has not been maintained.
Quality policy and objectives
The quality policy and quality objectives was/were not established by management with executive responsibility.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on FY2023. It asked for a Form 483.
Inspection ending Sep 2, 2022NAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-3382Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.20Management responsibility2 of 2
- 21 CFR 820.22Quality audit2 of 2
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.50Purchasing controls2 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
Show all 12
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.180General requirements1 of 2
- 21 CFR 820.181Device master record1 of 2
- 21 CFR 820.198Complaint files1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006111110 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006111110. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.