St. Jude Medical
387 Technology Cir Nw Ste 500, Atlanta, Georgia, United States
FEI 3004936110
FDA has 7 inspections on file for St. Jude Medical in Atlanta, Georgia (FEI 3004936110), from 2011 to 2024. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Jun 26, 2015. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 820.100(b) and 21 CFR 820.70(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 7, 2011–2024
- 3 OAI · 1 VAI · 3 NAI
- Inspections with 483 observations shown
- 3 of 7
- Form 483s posted by FDA
- None held for this FEI
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been established.
Periodic equipment inspections
Periodic inspections of equipment were not conducted to ensure adherence to applicable maintenance schedules.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Document review, approval by designated individual
Documents were not approved by designated individual(s) prior to issuance .
Show 4 more citations
Documentation
Corrective and preventive action activities and/or results have not been documented.
Document change records, maintained.
Records of changes to documents were not adequately maintained.
Production processes
Production processes were not developed, conducted and controlled to ensure that a device conforms to its specifications.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Quality audit - audit report review
Reports of results of quality audits are not reviewed by management having responsibility for the matters audited.
Documentation
Process validation activities have not been documented and approved.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Show 7 more citations
Production processes
Production processes were not developed, conducted, controlled, and monitored to ensure that a device conforms to its specifications.
Design history file
The design history file does not demonstrate that the design was developed following the approved design plan.
Lack of Written MDR Procedures
Written MDR procedures have not been maintained.
Lack of procedures, or not maintained
Document control procedures have not been maintained.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Equipment design and installation
Equipment used in the manufacturing process has not been appropriately installed to facilitate maintenance, adjustment, cleaning, and use.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch Monitoring3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Reasonably foreseeable risks or discomforts
The informed consent document lacked a description of reasonably foreseeable risks or discomforts to the subject.
Sponsors' general responsibilities
For an investigational study, investigators were not provided with the information they need to conduct an investigation properly, IRB review and approval were not ensured, reviewing IRBs were not promptly informed of significant new information about an investigation, and FDA was not promptly informed of significant new information about an investigation.
Procedures, identification of those which were experimental
The informed consent document did not contain a description of the procedures to be followed.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: DevicesCitations not publishedNot in FDA’s current Data Dashboard list (checked Nov 19, 2026)
- NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.40[Reserved]2 of 3
- 21 CFR 820.70Production and process controls2 of 3
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 50.25Elements of informed consent1 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 3
- 21 CFR 812.40General responsibilities of sponsors1 of 3
- 21 CFR 820.22Quality audit1 of 3
Show all 13
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.75Process validation1 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 3
- 21 CFR 820.90Nonconforming product1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Other FEIs recorded as St. Jude Medical
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
St. Jude MedicalFEI 2182269Minnetonka, Minnesota, United States
Last inspected 2019-03-20NAI
St. Jude MedicalFEI 3008308852Mahwah, New Jersey, United States
Last inspected 2010-04-19NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004936110 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004936110. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.