Skip to content

Suncoast Laboratories

6888 Lincoln Ave Ste G, Buena Park, California, United States

FEI 3004615749

FDA has 3 inspections on file for Suncoast Laboratories in Buena Park, California (FEI 3004615749), from 2008 to 2015. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.

The most recent OAI inspection ended on Dec 1, 2008. No warning letter, court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.200(a) and 21 CFR 820.22.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2008–2015
1 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
2 of 3
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    2. Investigation procedures

      The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not established, defined, and complete.

    3. Quality audit corrective action, reaudits {see also 820.100}

      Quality reaudits are not adequately performed.

    4. Servicing - Lack of or inadequate procedures

      Servicing procedures have not been established.

    5. Design validation - user needs and intended uses

      Design validation did not ensure that devices conform to defined user/patient needs and intended uses.

    Show 1 more citation
    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis was not performed.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices15 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality System Procedures

      Quality system procedures were not established.

    2. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not established.

    3. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    4. Design control - no procedures

      Procedures to control the design process of the device were not established, defined, documented, and implemented.

    5. Procedures

      The procedures addressing documentation of corrective and preventive action activities were not established.

    Show 10 more citations
    1. General

      Complaint handling procedures for complaints have not been established.

    2. DMR device specifications

      The device master record does not include or refer to the location of all, device, and software specifications.

    3. Software validation for automated processes

      Software used as part of production and the quality system has not been fully validated for its intended use according to an established protocol.

    4. Lack of procedures, or not maintained

      Document control procedures were not established.

    5. Quality policy and objectives

      The quality policy and objectives were not established.

    6. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not established.

    7. Specific non-conforming product procedures

      Procedures for addressing the identification, documentation, evaluation, and investigation of nonconforming product were not defined and documented.

    8. Lack of or inadequate DHR procedures

      Procedures have not been established to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.

    9. Management review - Lack of or inadequate procedures

      Procedures for management review are not established.

    10. Servicing - Lack of or inadequate procedures

      Servicing procedures have not been established.

Regulations FDA cited here

The 13 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.

Show all 13

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3004615749 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3004615749. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.