Suncoast Laboratories
6888 Lincoln Ave Ste G, Buena Park, California, United States
FEI 3004615749
FDA has 3 inspections on file for Suncoast Laboratories in Buena Park, California (FEI 3004615749), from 2008 to 2015. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.
The most recent OAI inspection ended on Dec 1, 2008. No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.200(a) and 21 CFR 820.22.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2008–2015
- 1 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Investigation procedures
The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not established, defined, and complete.
Quality audit corrective action, reaudits {see also 820.100}
Quality reaudits are not adequately performed.
Servicing - Lack of or inadequate procedures
Servicing procedures have not been established.
Design validation - user needs and intended uses
Design validation did not ensure that devices conform to defined user/patient needs and intended uses.
Show 1 more citation
Design Validation - Risk analysis not performed/inadequate
Risk analysis was not performed.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality System Procedures
Quality system procedures were not established.
Quality audits - Lack of or inadequate procedures
Procedures for conducting quality audits were not established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Design control - no procedures
Procedures to control the design process of the device were not established, defined, documented, and implemented.
Procedures
The procedures addressing documentation of corrective and preventive action activities were not established.
Show 10 more citations
General
Complaint handling procedures for complaints have not been established.
DMR device specifications
The device master record does not include or refer to the location of all, device, and software specifications.
Software validation for automated processes
Software used as part of production and the quality system has not been fully validated for its intended use according to an established protocol.
Lack of procedures, or not maintained
Document control procedures were not established.
Quality policy and objectives
The quality policy and objectives were not established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not established.
Specific non-conforming product procedures
Procedures for addressing the identification, documentation, evaluation, and investigation of nonconforming product were not defined and documented.
Lack of or inadequate DHR procedures
Procedures have not been established to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.
Management review - Lack of or inadequate procedures
Procedures for management review are not established.
Servicing - Lack of or inadequate procedures
Servicing procedures have not been established.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit2 of 2
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 820.200Servicing2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.40[Reserved]1 of 2
Show all 13
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.181Device master record1 of 2
- 21 CFR 820.184Device history record1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004615749 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004615749. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.