Skip to content

Qiagen Sciences LLC

19300 Germantown Rd, Germantown, Maryland, United States

FEI 3003572099

FDA has 7 inspections on file for Qiagen Sciences LLC in Germantown, Maryland (FEI 3003572099), from 2010 to 2021. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.

The most recent OAI inspection ended on Feb 5, 2016. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 820.250(b) and 21 CFR 820.30(i).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2010–2021
1 OAI · 3 VAI · 3 NAI
Inspections with 483 observations shown
4 of 7
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest FY2023
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    2. Sampling methods - Lack of or inadequate procedures

      Procedures to ensure sampling methods are adequate for their intended use have not been adequately established.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Project Evaluation: Devices15 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design output - Lack of or inadequate procedures

      Procedures for design output have not been adequately established.

    2. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been adequately established.

    3. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    4. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate and is incomplete .

    5. Design transfer - Lack of or inadequate procedures

      Procedures for design transfer have not been adequately established.

    Show 10 more citations
    1. Report of violation of the Act (see 803.52(e)(9))

      A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.

    2. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers and contractors have not been adequately established.

    3. Records of MDR Investigation

      Investigation records of MDR reportable complaints do not include required information. Specifically ***

    4. Records of complaint investigation

      Records of complaint investigations do not include required information.

    5. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

    6. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    7. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    8. Software validation for automated processes

      Software used as part of production has not been adequately validated for its intended use according to an established protocol.

    9. Sampling plans

      Sampling plans are not written and based on valid statistical rationale .

    10. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  6. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  7. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.

  • Inspection ending Jul 16, 2021VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. FY2023FDA-FOIA-2023-7530Form 483Closed
  • Inspection ending Apr 13, 2018

    1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.

    1. Sep 7, 2021FDA-FOIA-2021-5860Form 483No records found

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 7 inspections, with how many of them cited each.

Show all 10

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003572099 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003572099. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.