Bedfont Scientific, Ltd.
Station Road, Maidstone, United Kingdom
FEI 3002806497
FDA has 4 inspections on file for Bedfont Scientific, Ltd. in Maidstone, United Kingdom (FEI 3002806497), from 2011 to 2019. The latest ended on and was classified OAI, official action indicated. FDA has issued 2 warning letters under this FEI, the latest on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
5 import refusals are filed under this FEI, the latest on . FDA’s Import Alert 89-08 Red List names this firm at this address. FDA printed no FEI on the entry. No court action is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.100(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 4, 2011–2019
- 2 OAI · 1 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Nov 15, 2023
- Import refusals
- 5 refusals
- Import alerts
- 89-08Red List
- Matched on FDA’s firm name and street address
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design control - no procedures
Procedures for design control have not been established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Document change records, maintained.
Records of changes to documents were not maintained.
Documentation
Acceptance activities were not adequately documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Sampling plans
Sampling plans are not written and based on valid statistical rationale.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Training records
Personnel training is not documented.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained, and implemented.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Show 4 more citations
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Software validation for automated processes
Software used as part of production and the quality system has not been adequately validated for its intended use according to an established protocol.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Nov 15, 2023. Each one asked for a Form 483.
Inspection ending Oct 31, 2019OAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Nov 15, 2023FDA-FOIA-2023-10241Form 483Closed
- Oct 4, 2023FDA-FOIA-2023-8608Form 483· 1 of 8 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 3
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 820.22Quality audit1 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.40[Reserved]1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.70Production and process controls1 of 3
Show all 13
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 3
- 21 CFR 820.90Nonconforming product1 of 3
- 21 CFR 820.198Complaint files1 of 3
- 21 CFR 820.250Statistical techniques1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3002806497, one row per refused product line.
5 refusals from 2019-11-26 to 2026-01-13 · charged NO 510(K) on 5 · no FDA sample analysis
FDA firm profile- 2026-01-13
EQUIPMENT, LABORATORY, GENERAL PURPOSE, LABELED OR PROMOTED FOR A SPECIFIC MEDICAL USE75LXG
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-2208093-5/10002/1
- 2025-05-28
EQUIPMENT, LABORATORY, GENERAL PURPOSE, LABELED OR PROMOTED FOR A SPECIFIC MEDICAL USE75LXG
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-7113380-7/21/1
- 2023-02-10
SYSTEM, BREATH MEASUREMENT75NRH
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsSCS-5748050-4/11/1
- 2022-10-19
EQUIPMENT, LABORATORY, GENERAL PURPOSE, LABELED OR PROMOTED FOR A SPECIFIC MEDICAL USE75LXG
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northeast ImportsM50-0877518-6/31/1
- 2019-11-26
EQUIPMENT, LABORATORY, GENERAL PURPOSE, LABELED OR PROMOTED FOR A SPECIFIC MEDICAL USE75LXG
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsSCS-7322047-8/11/1
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 89-08Red ListDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Import alert pageAlert on fda.gov
Date published 2019-06-04
- FDA’s entry
- Bedfont Scientific, Ltd. · Station Road, Maidstone, Kent UNITED KINGDOM
- FDA’s inspection record for FEI 3002806497
- Bedfont Scientific, Ltd. · Station Road, Maidstone, United Kingdom
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
6 products
- 75 L - - XGEquipment, Laboratory, General Purpose, Labeled or Promoted for a Specific Medical Use
- 75 L - - XGEquipment, Laboratory, General Purpose, Labeled or Promoted for a Specific Medical Use
- 75 N - - RHSystem, Breath Measurement
Show 3 more products
- 82 M - - VMAutoantibodies, Endomysial (Tissue Transglutaminase)
- 89 I - - SAMassager, Therapeutic, Electric
- 89 N - - GXStimulator, Muscle, Powered, for Muscle Conditioning
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002806497 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002806497. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.