Skip to content

bioMerieux, Inc.

100 Rodolphe St, Durham, North Carolina, United States

FEI 3002769706

FDA has 4 inspections on file for bioMerieux, Inc. in Durham, North Carolina (FEI 3002769706), from 2012 to 2017. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request asked FDA for the 483 from a later inspection, ending Jul 1, 2024, that FDA's Data Dashboard doesn't list yet.

The most recent OAI inspection ended on Jun 14, 2013. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
4, 2012–2017
2 OAI · 0 VAI · 2 NAI
Inspections with 483 observations shown
2 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Jul 17, 2024
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Supplier oversight

      The type and extent of control to be exercised over the product, suppliers and contractors was not clearly defined.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices14 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    2. Lack of or inadequate procedures for handling

      Procedures for product handling have not been adequately established.

    3. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a process have not been adequately established.

    4. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated where necessary.

    5. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

    Show 9 more citations
    1. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    2. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    3. Personnel

      Personnel do not have the necessary training to perform their jobs.

    4. Periodic equipment inspections

      Periodic inspections of equipment were not conducted to ensure adherence to applicable maintenance schedules .

    5. Equipment control activity documentation

      Equipment maintenance activities have not been documented.

    6. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

    7. Training records

      Personnel training is not documented. Electronic records are used, but they do not meet requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.

    8. Procedures not adequately established or maintained

      Document control procedures have not been adequately established .

    9. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Jul 17, 2024. It asked for a Form 483.

  • Inspection ending Jul 1, 2024

    1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.

    1. Jul 17, 2024FDA-FOIA-2024-6268Form 483 + EIRClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.

Show all 11

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002769706 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002769706. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.