Skip to content

Pacific Consolidated Industries LLC

12201 Magnolia Ave, Riverside, California, United States

FEI 3002768081

FDA has 7 inspections on file for Pacific Consolidated Industries LLC in Riverside, California (FEI 3002768081), from 2008 to 2025. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.

The most recent OAI inspection ended on Feb 13, 2012. No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.100(b).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2008–2025
1 OAI · 3 VAI · 3 NAI
Inspections with 483 observations shown
6 of 7
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Aug 18, 2025
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    2. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities have not been documented, including analysis of sources of quality data, investigations of causes of nonconformities, the actions needed to correct or prevent recurrence of nonconforming product and other quality problems, the verification or validation of corrective actions, implementation of corrective and preventive actions, dissemination of information about quality problems or nonconforming product to responsible parties, and submission of information on quality problems and corrective and preventive actions for management review.

    2. Test and inspection data

      Service reports do not include applicable test and/or inspection data.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Process changes - review, evaluation and revalidation

      A validated process was not reviewed and evaluated and revalidated when changes or process deviations occurred.

    2. Design validation - documentation

      The results of design validation, including identification of the design, the date, and the individual(s) performing validation, were not adequately documented in the design history file.

    3. Design verification - documentation

      The design verification results, including identification of the design, method(s), and the individual(s) performing the verification, were not adequately documented in the design history file.

    4. Production processes

      Production processes were not controlled and monitored to ensure that a device conforms to its specifications.

    5. Incoming acceptance records, documentation

      Acceptance or rejection of incoming product was not documented.

    Show 4 more citations
    1. Change records, content

      Records of changes did not include the signature of the approving official(s), the approval date, and when the change became effective.

    2. Lack of or inadequate procedures for distribution

      Procedures for control and distribution of finished devices have not been established.

    3. Failure to Submit 5 Day Report

      A 5 day report was not submitted to FDA on Form 3500A within 5 workdays of becoming aware that a reportable MDR event necessitates remedial action to prevent an unreasonable risk of substantial harm to the public health.

    4. Complaints

      Complaint files are not adequately maintained.

  6. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  7. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities and the actions needed to correct or prevent recurrence of nonconforming product and other quality problems.

    2. Quality Audits - defined intervals

      Quality audits did not verify that the quality system is effective in fulfilling your quality system objectives.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Aug 18, 2025. It asked for a Form 483.

  • Inspection ending Jul 25, 2025VAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 16, 2025. FDAgent holds no copy that FDA posted.

    1. Aug 18, 2025FDA-FOIA-2025-9963Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 7 inspections, with how many of them cited each.

Show all 12

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002768081 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002768081. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.