Skip to content

The LifeStyle Company, Inc.

6 Paragon Way Ste 112, Freehold, New Jersey, United States

FEI 3002224304

FDA has 4 inspections on file for The LifeStyle Company, Inc. in Freehold, New Jersey (FEI 3002224304), from 2008 to 2017. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.198(c) and 21 CFR 820.80(b).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2008–2017
0 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
4 of 4
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Document review, approval by designated individual

      Documents were not approved by designated individual(s) prior to issuance .

    3. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    4. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated and investigated where necessary.

    5. Personnel

      Personnel do not have the necessary education, training and experience to perform their jobs.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been established.

    2. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    3. Quality audits - auditor independence

      Individuals who conduct quality audits have direct responsibility for the matters being audited.

    4. Documentation

      Process validation activities have not been adequately documented.

    5. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Acceptance activities were not documented.

    2. Documentation

      In-process inspections, tests, or other verification activities and approvals were not documented.

    3. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure that equipment is routinely calibrated were not established and implemented.

    2. Equipment suitability & capability

      Certain measuring and test equipment is not suitable for its intended purposes or capable of producing valid results.

    3. Procedure for analysis of data sources

      The corrective and preventive procedures addressing the analysis of sources of quality data to identify existing and potential causes of nonconforming product or other quality problems were not implemented.

    4. Lack of or inadequate final acceptance procedures

      Procedures for acceptance or rejection of finished device production runs, lots, or batches were not implemented.

    5. Storage procedures to avoid release of unsuitable product

      Procedures for controlling the storage of product in storage areas and stock rooms are not adequate to ensure that no obsolete, rejected, or deteriorated product is distributed.

    Show 3 more citations
    1. Process changes - review, evaluation and revalidation

      A validated process was not revalidated when changes or process deviations occurred.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance or rejection of incoming product were not implemented.

    3. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers were not sufficiently evaluated and selected on the basis of their ability to meet specified requirements.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.

Show all 12

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002224304 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002224304. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.