ATC Technologies, Inc.
30b Upton Dr, Wilmington, Massachusetts, United States
FEI 3001595430
FDA has 4 inspections on file for ATC Technologies, Inc. in Wilmington, Massachusetts (FEI 3001595430), from 2009 to 2019. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2009–2019
- 0 OAI · 4 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Equipment control activity documentation
Equipment calibrations, inspections, checks and maintenance activities have not been documented.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Acceptable supplier records, inadequate records
Records of acceptable suppliers and contractors have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Show 4 more citations
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Equipment suitability & capability
Certain inspection and measuring equipment is not suitable for its intended purposes.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design history file
The design history file does not demonstrate that the design was developed following the approved design plan and the design control requirements of 21 CFR 820.
Nature and details of complaint
Records of complaint investigations do not include the nature and details of the complaint.
Design validation- Lack of or inadequate procedures
Procedures for validating the device design were not implemented.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Documentation
Corrective and preventive action activities have not been documented, including the verification or validation of corrective actions, implementation of corrective and preventive actions, and dissemination of information about quality problems or nonconforming product to responsible parties.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action4 of 4
- 21 CFR 820.20Management responsibility2 of 4
- 21 CFR 820.50Purchasing controls2 of 4
- 21 CFR 820.70Production and process controls2 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 4
- 21 CFR 820.90Nonconforming product2 of 4
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 4
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 4
Show all 12
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.30Design controls1 of 4
- 21 CFR 820.75Process validation1 of 4
- 21 CFR 820.198Complaint files1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001595430 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001595430. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.