ATC Technologies, Inc.

30b Upton Dr, Wilmington, Massachusetts, United States

FEI 3001595430

FDA has 4 inspections on file for ATC Technologies, Inc. in Wilmington, Massachusetts (FEI 3001595430), from 2009 to 2019. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Four regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2009–2019
0 OAI · 4 VAI · 0 NAI
Inspections with 483 observations shown
4 of 4
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Equipment control activity documentation

      Equipment calibrations, inspections, checks and maintenance activities have not been documented.

    2. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    3. Acceptable supplier records, inadequate records

      Records of acceptable suppliers and contractors have not been adequately established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    4. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

    5. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    Show 4 more citations
    1. Procedures for product review,disposition lack of/inadequate

      Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    3. Equipment suitability & capability

      Certain inspection and measuring equipment is not suitable for its intended purposes.

    4. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.

    2. Product rework procedures, Lack of or inadequate procedures

      Procedures for rework of nonconforming product have not been adequately established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design history file

      The design history file does not demonstrate that the design was developed following the approved design plan and the design control requirements of 21 CFR 820.

    2. Nature and details of complaint

      Records of complaint investigations do not include the nature and details of the complaint.

    3. Design validation- Lack of or inadequate procedures

      Procedures for validating the device design were not implemented.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    5. Documentation

      Corrective and preventive action activities have not been documented, including the verification or validation of corrective actions, implementation of corrective and preventive actions, and dissemination of information about quality problems or nonconforming product to responsible parties.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.

Show all 12

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3001595430 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3001595430. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.