ConMed Corporation
525 French Rd, Utica, New York, United States
FEI 3001451571
FDA has 8 inspections on file for ConMed Corporation in Utica, New York (FEI 3001451571), from 2011 to 2022. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
3 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.198(a), 21 CFR 820.100(b) and 21 CFR 820.30(g).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 8, 2011–2022
- 0 OAI · 7 VAI · 1 NAI
- Inspections with 483 observations shown
- 6 of 8
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 3 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Reporting Information Reasonably Known
An MDR report submitted to FDA did not include all information that was reasonably known to the manufacturer.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Process changes - review, evaluation and revalidation
A validated process was not revalidated when changes or process deviations occurred.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Equipment control activity documentation
Equipment maintenance activities have not been documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Complaints
Complaint files are not adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Complaints
Complaint files are not adequately maintained.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files4 of 6
- 21 CFR 820.100Corrective and preventive action3 of 6
- 21 CFR 820.30Design controls2 of 6
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 6
- 21 CFR 820.50Purchasing controls1 of 6
- 21 CFR 820.70Production and process controls1 of 6
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 6
- 21 CFR 820.75Process validation1 of 6
Show all 9
- 21 CFR 820.90Nonconforming product1 of 6
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3001451571, one row per refused product line.
3 refusals on 2018-05-03 · charged FRNMFGREG on 3 · no FDA sample analysis
FDA firm profile- 2018-05-03
ARTHROSCOPE AND ACCESSORIES87HRX
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southwest ImportsUPS-7335464-0/61/1
- 2018-05-03
ARTHROSCOPE AND ACCESSORIES87HRX
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southwest ImportsUPS-7335464-0/41/1
- 2018-05-03
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT87LXH
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southwest ImportsUPS-7335464-0/11/1
Other FEIs recorded as ConMed Corporation
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
ConMed CorporationFEI 3004491689Greenwood Village, Colorado, United States
Last inspected 2026-04-06VAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001451571 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001451571. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.