Gynex Corporation
14603 NE 87th St, Redmond, Washington, United States
FEI 3001238683
FDA has 7 inspections on file for Gynex Corporation in Redmond, Washington (FEI 3001238683), from 2009 to 2020. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 7, 2009–2020
- 0 OAI · 5 VAI · 2 NAI
- Inspections with 483 observations shown
- 6 of 7
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Complaints
Complaint files are not adequately maintained.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers were not evaluated based on their ability to meet specified requirements.
Show 1 more citation
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Change records, content
Records of changes did not include a description of the change.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Show 1 more citation
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of procedures, or not maintained
Document control procedures have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Written record of complaint to include findings, follow-up
Written records of investigation of a drug complaint do not include the findings of the investigation and the follow-up.
Defined areas of adequate size for operations
The separate or defined areas necessary to prevent contamination or mix-ups are deficient.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Contract drug products--lack of responsibility
The quality control unit lacks responsibility for approving or rejecting drug products manufactured, processed, and packed under contract by another company.
Records fail to include
Records kept for each different labeling and packaging material shipment fail to include results of examination or testing and a statement of whether the shipment was accepted or rejected.
Complaint Handling Procedure
Procedures describing the handling of written and oral complaints related to drug products are deficiently written or followed.
Written procedures describing in detail
There is a lack of written procedures describing in sufficient detail the receipt, identification, storage, handling, sampling, and examination of labeling and packaging materials.
Returned drug procedures in writing and followed
Procedures describing the holding and testing of returned drug products are not in writing.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.50Purchasing controls3 of 6
- 21 CFR 211.198Complaint files2 of 6
- 21 CFR 820.40[Reserved]2 of 6
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 6
- 21 CFR 820.100Corrective and preventive action2 of 6
- 21 CFR 820.198Complaint files2 of 6
- 21 CFR 211.22Responsibilities of quality control unit1 of 6
- 21 CFR 211.42Design and construction features1 of 6
Show all 14
- 21 CFR 211.122Materials examination and usage criteria1 of 6
- 21 CFR 211.204Returned drug products1 of 6
- 21 CFR 820.25Personnel1 of 6
- 21 CFR 820.30Design controls1 of 6
- 21 CFR 820.70Production and process controls1 of 6
- 21 CFR 820.90Nonconforming product1 of 6
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001238683 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001238683. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.