Klarity Medical Products LLC
600 Industrial Pkwy, Heath, Ohio, United States
FEI 3001236129
FDA has 7 inspections on file for Klarity Medical Products LLC in Heath, Ohio (FEI 3001236129), from 2010 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 4 citations from this inspection. They are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.
The most recent OAI inspection ended on Feb 25, 2020. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 7, 2010–2026
- 1 OAI · 3 VAI · 3 NAI
- Inspections with 483 observations shown
- 7 of 7
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest FY2023
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 8.3.4
Reworked product not verified
Reworked product was not adequately verified to ensure it meets applicable acceptance criteria and regulatory requirements.
- ISO 13485:2016 Clause 7.5.6
Procedures for validation of processes not documented
Procedures for validation of processes were not adequately documented.
- ISO 13485:2016 Clause 7.3.9
Records of design and development changes and their reviews not maintained
Records of design and development changes and necessary actions associated with design and development changes were not maintained.
- ISO 13485:2016 Clause 7.4.1
Records of supplier (re)evaluation, selection, and monitoring not maintained
Records of the results of monitoring of supplier capability or performance were not maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DMR - not or inadequately maintained
A device master record has not been maintained.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
Lack of procedures, or not maintained
Document control procedures have not been established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been established.
Show 2 more citations
DMR - not or inadequately maintained
A device master record has not been maintained.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been established.
Supplier notification of changes
There is no agreement with suppliers, contractors and consultants to notify you of changes in the product or service.
Show 2 more citations
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documented evaluation
The evaluation of potential suppliers, contractors, and consultants was not documented.
Processing MDRs {see also 803, 804}
Complaint handling procedures have not been implemented to ensure that all complaints are evaluated to determine whether the complaint should be filed as a Medical Device Report.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.
Inspection ending May 20, 2026NAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 2, 2026. FDAgent holds no copy that FDA posted.
- Aug 21, 2026FDA-FOIA-2026-8127Form 483· 1 of 4 sitesOpen
- Aug 4, 2026FDA-FOIA-2026-7501Form 483· 1 of 5 sitesRecords released
Inspection ending Feb 16, 2023VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-4199Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.50Purchasing controls5 of 7
- 21 CFR 820.100Corrective and preventive action4 of 7
- 21 CFR 820.80Receiving, in-process, and finished device acceptance3 of 7
- 21 CFR 820.181Device master record2 of 7
- 21 CFR 820.198Complaint files2 of 7
- 21 CFR 820.25Personnel1 of 7
- 21 CFR 820.30Design controls1 of 7
- 21 CFR 820.40[Reserved]1 of 7
Show all 12
- 21 CFR 820.70Production and process controls1 of 7
- 21 CFR 820.75Process validation1 of 7
- 21 CFR 820.90Nonconforming product1 of 7
- 21 CFR 820.184Device history record1 of 7
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001236129 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001236129. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.