Laser Engineering, Inc.
475 Metroplex Dr Ste 401, Nashville, Tennessee, United States
FEI 3000204219
FDA has 5 inspections on file for Laser Engineering, Inc. in Nashville, Tennessee (FEI 3000204219), from 2008 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Apr 4, 2014. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.
12 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2008–2025
- 3 OAI · 0 VAI · 2 NAI
- Inspections with 483 observations shown
- 3 of 5
- Form 483s posted by FDA
- None held for this FEI
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety Act16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Safety interlock not provided
Each laser product was not provided with at least one safety interlock for each portion of the protective housing which is designed to be removed or displaced during operation or maintenance, when such removal or displacement permitted human access to laser or collateral radiation in excess of the accessible emission limits of Class I and table VI.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Acceptable supplier records, inadequate records
Records of acceptable suppliers and contractors have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Show 11 more citations
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Failure to submit
You did not submit a required supplemental report, prior to the introduction of the product into commerce, for a new or modified model within a model or chassis family.
Design history file
The design history file does not demonstrate that the design was developed following the approved design plan and does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Remedial action - documentation
Evaluations of out-of-calibration equipment and remedial actions taken were not documented.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been established.
Document review, approval by designated individual
Documents were not reviewed and not approved by designated individual(s) prior to issuance .
User information, adequate instructions
Adequate instructions for each laser product were not provided or caused to be provided for assembly, operation, and maintenance.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices19 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Remedial action - documentation
Evaluations of out-of-calibration equipment and remedial actions taken were not documented.
Aperture label - laser radiation
The warning statement "AVOID EXPOSURE---Laser radiation is emitted from this aperture" was not affixed in close proximity to each aperture through which is emitted accessible laser or collateral radiation in excess of the accessible emission limits of Class I and table VI of Section 1040.10(d).
50 mm classification warning
The following warning was not included in the user information: "CAUTION --- The use of optical instruments with this product will increase eye hazard", when use of a 50 millimeter diameter aperture stop in a laser product classified with a 7 millimeter diameter aperture stop resulted in a higher classification of the product.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
Show 14 more citations
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of required elements
An annual report which you submitted did not summarize the contents of the records required by 21 CFR 1002.30(a) and provide the volume of products produced, sold, or installed.
Protective housing
Each laser product is not provided with a protective housing that prevents human access during operations to laser and collateral radiation that exceeds the limits of Class I and table VI, respectively, wherever and whenever such human access is not necessary for the product to perform its intended function(s).
DIPH label wording - Class IV
Labels regarding Class IV accessible laser radiation which are in close proximity to the opening created by the removal or displacement of such portion of the protective housing were not provided for defeatably interlocked protective housings.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Reproductions of warnings
Adequate instructions for service adjustments and service procedures were not provided or caused to be provided for each laser model, including legible reproductions of required labels and hazard warnings.
Supplier notification of changes
There is no agreement with suppliers, contractors, and consultants to notify you of changes in the product or service.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Acceptable supplier records, inadequate records
Records of acceptable suppliers and contractors have not been established.
Document review, approval by designated individual
Documents were not reviewed and not approved by designated individual(s) prior to issuance .
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Safety interlock not provided
Each laser product was not provided with at least one safety interlock for each portion of the protective housing which is designed to be removed or displaced during operation or maintenance, when such removal or displacement permitted human access to laser or collateral radiation in excess of the accessible emission limits of Class I and table VI.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety Act10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training
Employees have not been adequately trained.
No Service reports
Service reports were not documented.
Analysis of data sources
Appropriate sources of quality data are not adequately analyzed to identify existing and potential causes of nonconforming product and other quality problems.
Identifying relevant info for management review
Procedures addressing identification of relevant information to be submitted for management review were not implemented.
Failure to submit
You did not submit an annual report for products requiring one.
Show 5 more citations
Quality Audits - defined intervals
Quality audits did not verify that the quality system is effective in fulfilling your quality system objectives.
General
Complaint handling procedures for complaints have not been implemented.
Document review, approval documentation
The documentation of approval of documents does not include the signature of the approving official.
Training records
Employee training is not fully documented.
Management review - Lack of or inadequate procedures
Procedures for management review are not established, defined, and documented.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.20Management responsibility3 of 3
- 21 CFR 820.22Quality audit3 of 3
- 21 CFR 820.40[Reserved]3 of 3
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 820.198Complaint files3 of 3
- 21 CFR 820.30Design controls2 of 3
- 21 CFR 820.50Purchasing controls2 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 3
Show all 16
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 3
- 21 CFR 820.184Device history record2 of 3
- 21 CFR 1002.13Annual reports2 of 3
- 21 CFR 1040.10Laser products2 of 3
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.200Servicing1 of 3
- 21 CFR 1002.11Supplemental reports1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3000204219 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3000204219. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.