Laser Engineering, Inc.

475 Metroplex Dr Ste 401, Nashville, Tennessee, United States

FEI 3000204219

FDA has 5 inspections on file for Laser Engineering, Inc. in Nashville, Tennessee (FEI 3000204219), from 2008 to 2025. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Apr 4, 2014. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.

12 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
5, 2008–2025
3 OAI · 0 VAI · 2 NAI
Inspections with 483 observations shown
3 of 5
Form 483s posted by FDA
None held for this FEI
Warning letters
· CDRH ↗
2 on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety Act16 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    2. Safety interlock not provided

      Each laser product was not provided with at least one safety interlock for each portion of the protective housing which is designed to be removed or displaced during operation or maintenance, when such removal or displacement permitted human access to laser or collateral radiation in excess of the accessible emission limits of Class I and table VI.

    3. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    4. Acceptable supplier records, inadequate records

      Records of acceptable suppliers and contractors have not been adequately established.

    5. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    Show 11 more citations
    1. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    3. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    4. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    5. Failure to submit

      You did not submit a required supplemental report, prior to the introduction of the product into commerce, for a new or modified model within a model or chassis family.

    6. Design history file

      The design history file does not demonstrate that the design was developed following the approved design plan and does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    7. Remedial action - documentation

      Evaluations of out-of-calibration equipment and remedial actions taken were not documented.

    8. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    9. Design transfer - Lack of or inadequate procedures

      Procedures for design transfer have not been established.

    10. Document review, approval by designated individual

      Documents were not reviewed and not approved by designated individual(s) prior to issuance .

    11. User information, adequate instructions

      Adequate instructions for each laser product were not provided or caused to be provided for assembly, operation, and maintenance.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices19 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    2. Remedial action - documentation

      Evaluations of out-of-calibration equipment and remedial actions taken were not documented.

    3. Aperture label - laser radiation

      The warning statement "AVOID EXPOSURE---Laser radiation is emitted from this aperture" was not affixed in close proximity to each aperture through which is emitted accessible laser or collateral radiation in excess of the accessible emission limits of Class I and table VI of Section 1040.10(d).

    4. 50 mm classification warning

      The following warning was not included in the user information: "CAUTION --- The use of optical instruments with this product will increase eye hazard", when use of a 50 millimeter diameter aperture stop in a laser product classified with a 7 millimeter diameter aperture stop resulted in a higher classification of the product.

    5. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    Show 14 more citations
    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Lack of required elements

      An annual report which you submitted did not summarize the contents of the records required by 21 CFR 1002.30(a) and provide the volume of products produced, sold, or installed.

    3. Protective housing

      Each laser product is not provided with a protective housing that prevents human access during operations to laser and collateral radiation that exceeds the limits of Class I and table VI, respectively, wherever and whenever such human access is not necessary for the product to perform its intended function(s).

    4. DIPH label wording - Class IV

      Labels regarding Class IV accessible laser radiation which are in close proximity to the opening created by the removal or displacement of such portion of the protective housing were not provided for defeatably interlocked protective housings.

    5. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    6. Reproductions of warnings

      Adequate instructions for service adjustments and service procedures were not provided or caused to be provided for each laser model, including legible reproductions of required labels and hazard warnings.

    7. Supplier notification of changes

      There is no agreement with suppliers, contractors, and consultants to notify you of changes in the product or service.

    8. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    9. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    10. Acceptable supplier records, inadequate records

      Records of acceptable suppliers and contractors have not been established.

    11. Document review, approval by designated individual

      Documents were not reviewed and not approved by designated individual(s) prior to issuance .

    12. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    13. Safety interlock not provided

      Each laser product was not provided with at least one safety interlock for each portion of the protective housing which is designed to be removed or displaced during operation or maintenance, when such removal or displacement permitted human access to laser or collateral radiation in excess of the accessible emission limits of Class I and table VI.

    14. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety Act10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training

      Employees have not been adequately trained.

    2. No Service reports

      Service reports were not documented.

    3. Analysis of data sources

      Appropriate sources of quality data are not adequately analyzed to identify existing and potential causes of nonconforming product and other quality problems.

    4. Identifying relevant info for management review

      Procedures addressing identification of relevant information to be submitted for management review were not implemented.

    5. Failure to submit

      You did not submit an annual report for products requiring one.

    Show 5 more citations
    1. Quality Audits - defined intervals

      Quality audits did not verify that the quality system is effective in fulfilling your quality system objectives.

    2. General

      Complaint handling procedures for complaints have not been implemented.

    3. Document review, approval documentation

      The documentation of approval of documents does not include the signature of the approving official.

    4. Training records

      Employee training is not fully documented.

    5. Management review - Lack of or inadequate procedures

      Procedures for management review are not established, defined, and documented.

Regulations FDA cited here

The 16 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.

Show all 16

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3000204219 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3000204219. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.