FDA has 3 inspections on file for Sgarlato Med, LLC in San Jose, California (FEI 2939320), from 2010 to 2014. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 15 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Nine regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
3, 2010–2014
1 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices15 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Quality policy and objectives

      The quality objectives was/were not established by management with executive responsibility.

    4. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.

    5. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    Show 10 more citations
    1. Design history file

      The design history file was not established, does not demonstrate that the design was developed following the approved design plan, and does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    2. DMR - not or inadequately maintained

      A device master record has not been maintained.

    3. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    4. Acceptable supplier records, inadequate records

      Records of acceptable contractors have not been adequately established.

    5. Lack of or inadequate procedures for distribution

      Procedures for control and distribution of finished devices have not been adequately established.

    6. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    7. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    8. Personnel

      Personnel do not have the necessary experience to perform their jobs.

    9. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    10. Lack of or inadequate DHR procedures

      Procedures for device history records have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Acceptance activities were not adequately documented.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    4. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    5. Design validation - simulated testing

      The design was not validated under actual or simulated use conditions.

    Show 7 more citations
    1. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Sampling plans

      Sampling plans are not based on valid statistical rationale.

    4. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    5. Design control - no procedures

      Procedures for design control have not been established.

    6. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    7. Report of violation of the Act (see 803.52(e)(9))

      A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices18 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures were not followed for the identification, documentation, validation or verification, review, and approval of design changes before their implementation.

    2. Documentation

      Corrective and preventive action activities have not been documented, including the actions needed to correct or prevent recurrence of nonconforming product and other quality problems, the verification or validation of corrective actions, and implementation of corrective and preventive actions.

    3. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    4. Design input - documentation

      Design input requirements were not fully documented.

    5. Design verification procedures

      Procedures for verifying that design output meets design input were not implemented.

    Show 13 more citations
    1. Design review - Lack of or inadequate procedures

      Procedures for planning and conducting reviews of the design results at appropriate stages of the device's design development were not implemented.

    2. Report of violation of the Act (see 803.52(e)(9))

      A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.

    3. Design changes - validation vs. verification

      Procedures were not established to address when verification of design changes is sufficient in lieu of validation prior to their implementation.

    4. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not complete.

    5. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities were not complete and implemented.

    6. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not implemented.

    7. Investigation of device failures

      Complaints involving the possible failure of labeling to meet any of its specifications were not investigated where necessary.

    8. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    9. Quality Audit/Reaudit - conducted

      Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.

    10. Design control - no procedures

      Procedures to control the design process of the device were not implemented.

    11. Documented evaluation

      The evaluation of potential suppliers, contractors, and consultants was not documented.

    12. Design reviews performed following schedule

      Design reviews were not performed at appropriate times, following the review schedule.

    13. Processing MDRs {see also 803, 804}

      Complaint handling procedures have not been implemented to ensure that all complaints are evaluated to determine whether the complaint should be filed as a Medical Device Report.

Regulations FDA cited here

The 15 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Show all 15

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2939320 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2939320. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.