Sgarlato Med, LLC
1975 Hamilton Ave Ste 38, San Jose, California, United States
FEI 2939320
FDA has 3 inspections on file for Sgarlato Med, LLC in San Jose, California (FEI 2939320), from 2010 to 2014. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 15 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Nine regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 3, 2010–2014
- 1 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Quality policy and objectives
The quality objectives was/were not established by management with executive responsibility.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Show 10 more citations
Design history file
The design history file was not established, does not demonstrate that the design was developed following the approved design plan, and does not demonstrate that the design was developed following the requirements of 21 CFR 820.
DMR - not or inadequately maintained
A device master record has not been maintained.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Acceptable supplier records, inadequate records
Records of acceptable contractors have not been adequately established.
Lack of or inadequate procedures for distribution
Procedures for control and distribution of finished devices have not been adequately established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Personnel
Personnel do not have the necessary experience to perform their jobs.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Acceptance activities were not adequately documented.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Design validation - simulated testing
The design was not validated under actual or simulated use conditions.
Show 7 more citations
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Sampling plans
Sampling plans are not based on valid statistical rationale.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Design control - no procedures
Procedures for design control have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Report of violation of the Act (see 803.52(e)(9))
A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices18 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures were not followed for the identification, documentation, validation or verification, review, and approval of design changes before their implementation.
Documentation
Corrective and preventive action activities have not been documented, including the actions needed to correct or prevent recurrence of nonconforming product and other quality problems, the verification or validation of corrective actions, and implementation of corrective and preventive actions.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Design input - documentation
Design input requirements were not fully documented.
Design verification procedures
Procedures for verifying that design output meets design input were not implemented.
Show 13 more citations
Design review - Lack of or inadequate procedures
Procedures for planning and conducting reviews of the design results at appropriate stages of the device's design development were not implemented.
Report of violation of the Act (see 803.52(e)(9))
A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.
Design changes - validation vs. verification
Procedures were not established to address when verification of design changes is sufficient in lieu of validation prior to their implementation.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not complete.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities were not complete and implemented.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not implemented.
Investigation of device failures
Complaints involving the possible failure of labeling to meet any of its specifications were not investigated where necessary.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Quality Audit/Reaudit - conducted
Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.
Design control - no procedures
Procedures to control the design process of the device were not implemented.
Documented evaluation
The evaluation of potential suppliers, contractors, and consultants was not documented.
Design reviews performed following schedule
Design reviews were not performed at appropriate times, following the review schedule.
Processing MDRs {see also 803, 804}
Complaint handling procedures have not been implemented to ensure that all complaints are evaluated to determine whether the complaint should be filed as a Medical Device Report.
Regulations FDA cited here
The 15 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit3 of 3
- 21 CFR 820.30Design controls3 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance3 of 3
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 806.10Reports of corrections and removals2 of 3
- 21 CFR 820.25Personnel2 of 3
- 21 CFR 820.50Purchasing controls2 of 3
- 21 CFR 820.90Nonconforming product2 of 3
Show all 15
- 21 CFR 820.198Complaint files2 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.40[Reserved]1 of 3
- 21 CFR 820.160Distribution1 of 3
- 21 CFR 820.181Device master record1 of 3
- 21 CFR 820.184Device history record1 of 3
- 21 CFR 820.250Statistical techniques1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2939320 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2939320. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.