Busse Hospital Disposables, Inc.
75 Arkay Dr, Hauppauge, New York, United States
FEI 2433012
FDA has 6 inspections on file for Busse Hospital Disposables, Inc. in Hauppauge, New York (FEI 2433012), from 2008 to 2018. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.90(a) and 21 CFR 820.70(c).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2008–2018
- 0 OAI · 3 VAI · 3 NAI
- Inspections with 483 observations shown
- 4 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers and contractors have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR - not or inadequately maintained
A device history record has not been maintained.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not fully followed.
Storage under appropriate conditions
Drug products are not stored under appropriate conditions of temperature so that their identity, strength, quality, and purity are not affected.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Backup data not assured as exact and complete
Backup data is not assured as through keeping hard copy or alternate systems. Electronic records are used, but they do not meet audit trail requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.
Show 1 more citation
QSR
The quality system record has not been maintained. Electronic records are used, but they do not meet retention, system access limitation, systems documentation control and open systems control requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Submission of Supplement to Individual Report Within One Month
A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.70Production and process controls2 of 4
- 21 CFR 820.90Nonconforming product2 of 4
- 21 CFR 211.22Responsibilities of quality control unit1 of 4
- 21 CFR 211.68Automatic, mechanical, and electronic equipment1 of 4
- 21 CFR 211.142Warehousing procedures1 of 4
- 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 4
- 21 CFR 820.50Purchasing controls1 of 4
- 21 CFR 820.75Process validation1 of 4
Show all 12
- 21 CFR 820.100Corrective and preventive action1 of 4
- 21 CFR 820.150Storage1 of 4
- 21 CFR 820.184Device history record1 of 4
- 21 CFR 820.186Quality system record1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2433012 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2433012. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.