Biodex Medical Systems, Inc.
49 Natcon Dr, Shirley, New York, United States
FEI 2431314
FDA has 6 inspections on file for Biodex Medical Systems, Inc. in Shirley, New York (FEI 2431314), from 2011 to 2023. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.184, 21 CFR 820.40(b) and 21 CFR 803.50(a)(1).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2011–2023
- 0 OAI · 5 VAI · 1 NAI
- Inspections with 483 observations shown
- 5 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Document change records, maintained.
Records of changes to documents were not adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality control procedures
You have not established records containing a description of quality control procedures with respect to electronic product radiation safety.
DHR - not or inadequately maintained
A device history record has not been maintained.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been established.
Show 2 more citations
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by contractors and consultants have not been established.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Process validation results have not been adequately documented.
Traceability Lack of or inadequate {see also 820.120(e)}
Procedures for identifying with a control number each unit, lot, or batch of implantable or life supporting, life sustaining finished devices or appropriate components have not been established.
Examination for accuracy
Labeling was not sufficiently examined by a designated individual for accuracy including control numbers before release.
Production processes
Production processes were not controlled to ensure that a device conforms to its specifications.
DHR - not or inadequately maintained
A device history record has not been maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Change records, content
Records of changes did not include a description of the change.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Results of evaluation not in mfr's MDR file
The results of the evaluation of each event are not documented and maintained in the manufacturer's MDR event file.
DHR - not or inadequately maintained
A device history record has not been maintained.
Adverse events--all info not in file
MDR event files do not contain or reference all adverse event information in the possession of the reporting entity, including documentation of the deliberations and decision making process used to determine if an event was or was not reportable.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.184Device history record4 of 5
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 5
- 21 CFR 820.40[Reserved]2 of 5
- 21 CFR 820.70Production and process controls2 of 5
- 21 CFR 803.18What are the requirements for establishing and maintaining MDR files or records that apply to me?1 of 5
- 21 CFR 820.50Purchasing controls1 of 5
- 21 CFR 820.75Process validation1 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 5
Show all 11
- 21 CFR 820.120Device labeling1 of 5
- 21 CFR 1002.30Records to be maintained by manufacturers1 of 5
- 21 CFR 820.65Traceability1 of 5
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2431314 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2431314. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.