Skip to content

Biodex Medical Systems, Inc.

49 Natcon Dr, Shirley, New York, United States

FEI 2431314

FDA has 6 inspections on file for Biodex Medical Systems, Inc. in Shirley, New York (FEI 2431314), from 2011 to 2023. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.184, 21 CFR 820.40(b) and 21 CFR 803.50(a)(1).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
6, 2011–2023
0 OAI · 5 VAI · 1 NAI
Inspections with 483 observations shown
5 of 6
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    2. Document change records, maintained.

      Records of changes to documents were not adequately maintained.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality control procedures

      You have not established records containing a description of quality control procedures with respect to electronic product radiation safety.

    2. DHR - not or inadequately maintained

      A device history record has not been maintained.

    3. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    4. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    5. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been established.

    Show 2 more citations
    1. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by contractors and consultants have not been established.

    2. Incoming acceptance records, documentation

      Acceptance or rejection of incoming product was not documented.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Process validation results have not been adequately documented.

    2. Traceability Lack of or inadequate {see also 820.120(e)}

      Procedures for identifying with a control number each unit, lot, or batch of implantable or life supporting, life sustaining finished devices or appropriate components have not been established.

    3. Examination for accuracy

      Labeling was not sufficiently examined by a designated individual for accuracy including control numbers before release.

    4. Production processes

      Production processes were not controlled to ensure that a device conforms to its specifications.

    5. DHR - not or inadequately maintained

      A device history record has not been maintained.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Change records, content

      Records of changes did not include a description of the change.

    2. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Results of evaluation not in mfr's MDR file

      The results of the evaluation of each event are not documented and maintained in the manufacturer's MDR event file.

    2. DHR - not or inadequately maintained

      A device history record has not been maintained.

    3. Adverse events--all info not in file

      MDR event files do not contain or reference all adverse event information in the possession of the reporting entity, including documentation of the deliberations and decision making process used to determine if an event was or was not reportable.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 6 inspections, with how many of them cited each.

Show all 11

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2431314 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2431314. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.