21 CFR 820.65: Traceability
FDA's inspection data lists 6 Form 483 citations of 21 CFR 820.65 at 6 facilities, from November 2010 to June 27, 2019. eCFR shows this section removed on February 4, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 6
- At 6 inspections
- Facilities cited
- 6
- Distinct FEIs
- Most recent citation
- Jun 27, 2019
- Ad-Tech Medical Instrument Corporation
- Warning letters
- 0
- State a violation of it
The regulation
Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).
§ 820.65 Traceability.
Each manufacturer of a device that is intended for surgical implant into the body or to support or sustain life and whose failure to perform when properly used in accordance with instructions for use provided in the labeling can be reasonably expected to result in a significant injury to the user shall establish and maintain procedures for identifying with a control number each unit, lot, or batch of finished devices and where appropriate components. The procedures shall facilitate corrective action. Such identification shall be documented in the DHR.
21 CFR Part 820: Quality Management System Regulation, Subpart FRead on eCFR
Citations by year
Form 483 citations of 21 CFR 820.65 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 2 in 2015
| Year | Form 483 citations |
|---|---|
| 2010 | 1 |
| 2013 | 1 |
| 2014 | 1 |
| 2015 | 2 |
| 2019 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 6 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Traceability Lack of or inadequate {see also 820.120(e)}6 citations
Citation text
- “Procedures for identifying with a control number each unit, lot, or batch of implantable or life supporting, life sustaining finished devices or appropriate components have not been adequately established.”5 citations
- “Procedures for identifying with a control number each unit, lot, or batch of implantable or life supporting, life sustaining finished devices or appropriate components have not been established.”1 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 820.65, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Ad-Tech Medical Instrument Corporation2019-06-27 · FEI 2183456Traceability Lack of or inadequate {see also 820.120(e)}
- Buffalo Dental Manufacturing Co., Inc.2015-12-22 · FEI 2411056Traceability Lack of or inadequate {see also 820.120(e)}
- Biodex Medical Systems, Inc.2015-03-17 · FEI 2431314Traceability Lack of or inadequate {see also 820.120(e)}
- Philips North America, LLC2013-12-12 · FEI 1218950Traceability Lack of or inadequate {see also 820.120(e)}
Often cited at the same inspection
Of the 6 inspections where FDA cited 21 CFR 820.65, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action4 (67%)
- 21 CFR 820.184Device history record4 (67%)
- 21 CFR 820.198Complaint files3 (50%)
- 21 CFR 820.80Receiving, in-process, and finished device acceptance3 (50%)
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 (33%)
- 21 CFR 820.120Device labeling2 (33%)
Ask FDAgent about 21 CFR 820.65
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.65; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.