Hill-Rom Inc
1069 State Road 46 E, Batesville, Indiana, United States
FEI 1824206
FDA has 8 inspections on file for Hill-Rom Inc in Batesville, Indiana (FEI 1824206), from 2009 to 2024. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Feb 13, 2014. FDA issued a warning letter on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 8, 2009–2024
- 3 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 6 of 8
- 1 posted 483 not matched to an inspection
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 2 requests, latest Apr 7, 2025
- Warning letters
- · CDRH ↗
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 3 more citations
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
Identification procedures, Lack of or inadequate procedures
Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Show 1 more citation
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Acceptance status
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Identifying corrective & preventive actions
The procedures addressing identification of corrective and preventive actions were not complete.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Apr 7, 2025.
Inspection ending Nov 19, 2021VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Dec 6, 2023FDA-FOIA-2023-10887Form 483Closed
Inspection date not given
1 request. None of FDA’s log entries for these requests say records were released.
- Apr 7, 2025FDA-FOIA-2025-3232EIRDelogged (not a FOIA Request)
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files4 of 6
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?3 of 6
- 21 CFR 820.30Design controls2 of 6
- 21 CFR 820.90Nonconforming product2 of 6
- 21 CFR 820.100Corrective and preventive action2 of 6
- 21 CFR 820.60Identification1 of 6
- 21 CFR 820.70Production and process controls1 of 6
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 6
Show all 9
- 21 CFR 820.86Acceptance status1 of 6
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1824206, one row per refused product line.
1 refusal on 2020-05-16 · charged NOT LISTED on 1, FRNMFGREG on 1 · no FDA sample analysis
FDA firm profile- 2020-05-16
BED, AC-POWERED ADJUSTABLE HOSPITAL80FNL
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports147-0183390-8/11/2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1824206 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1824206. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.