Skip to content

Hill-Rom, Inc. Form 483, January 2013

FDA issued this Form 483 to Hill-Rom, Inc. in Batesville, Indiana (FEI 1824206) at the end of a device inspection that ended January 11, 2013. FDA classified the inspection Official Action Indicated (OAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Jan 11, 2013
Devices · Compliance: Devices
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Jan 17, 2024
4,023 days after the inspection
Facility
FEI 1824206
Batesville, Indiana

FDA's 6 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Design validation- Lack of or inadequate procedures

    Procedures for design validation have not been adequately established.

  2. Design verification - Lack of or inadequate procedures

    Procedures for design verification have not been adequately established.

  3. Identification procedures, Lack of or inadequate procedures

    Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been adequately established.

  4. Investigation of device failures

    Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.

  5. Design verification - output does not meet input requirement

    Design verification does not confirm that design output meets design input requirements.

  6. Lack of Written MDR Procedures

    Written MDR procedures have not been implemented.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.