Edwards Lifesciences, LLC
12050 S Lone Peak Pkwy, Draper, Utah, United States
FEI 1713910
FDA has 8 inspections on file for Edwards Lifesciences, LLC in Draper, Utah (FEI 1713910), from 2009 to 2021. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Apr 11, 2014. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 820.80(d) and 21 CFR 820.90(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 8, 2009–2021
- 1 OAI · 1 VAI · 6 NAI
- Inspections with 483 observations shown
- 3 of 8
- 1 posted 483 not matched to an inspection
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 1 request, latest Jan 14, 2025
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation of validated process performance
There is inadequate documentation of monitoring and control methods and data for a validated process.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record .
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Project Evaluation: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate and is incomplete.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 5 more citations
Packaging
Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Jan 14, 2025. It asked for a Form 483.
Inspection ending Apr 11, 2014OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Jan 14, 2025FDA-FOIA-2025-420Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 820.75Process validation2 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 3
- 21 CFR 820.90Nonconforming product2 of 3
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.70Production and process controls1 of 3
Show all 11
- 21 CFR 820.130Device packaging1 of 3
- 21 CFR 820.184Device history record1 of 3
- 21 CFR 820.198Complaint files1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Other FEIs recorded as Edwards Lifesciences, LLC
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Edwards Lifesciences, LLCFEI 2015691Irvine, California, United States
Last inspected 2024-08-15NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1713910 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1713910. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.