Sonogage Inc
26650 Renaissance Pkwy Ste 7, Cleveland, Ohio, United States
FEI 1526843
FDA has 5 inspections on file for Sonogage Inc in Cleveland, Ohio (FEI 1526843), from 2009 to 2023. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
Seven regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2009–2023
- 0 OAI · 3 VAI · 2 NAI
- Inspections with 483 observations shown
- 4 of 5
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Process validation activities and results have not been adequately documented and adequately approved.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of procedures, or not maintained
Document control procedures have not been maintained.
Show 2 more citations
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training records
Personnel training is not documented.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of procedures, or not maintained
Document control procedures have not been established.
Show 5 more citations
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Documentation
Process validation activities and results have not been adequately documented and adequately approved.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Acceptance records
The device history record does not include acceptance records that demonstrate the device is manufactured in accordance with the device master record.
Specific non-conforming product procedures
Procedures for addressing the identification, documentation, evaluation, segregation, disposition, and investigation of nonconforming product were not defined.
Incoming inspection, testing, verification
Incoming product was not adequately inspected or tested to verify conformance to specifications.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not established.
Procedure for analysis of data sources
The corrective and preventive procedures addressing the analysis of sources of quality data to identify existing and potential causes of nonconforming product or other quality problems were not complete.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.50Purchasing controls3 of 4
- 21 CFR 820.90Nonconforming product3 of 4
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 820.22Quality audit2 of 4
- 21 CFR 820.40[Reserved]2 of 4
- 21 CFR 820.75Process validation2 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 4
- 21 CFR 820.198Complaint files2 of 4
Show all 12
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.70Production and process controls1 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 4
- 21 CFR 820.184Device history record1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1526843 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1526843. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.