Multi-Med, Inc.

26 Victoria Ct, Keene, New Hampshire, United States

FEI 1221765

FDA has 7 inspections on file for Multi-Med, Inc. in Keene, New Hampshire (FEI 1221765), from 2009 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

The most recent OAI inspection ended on Sep 11, 2009. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.40.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2009–2024
1 OAI · 3 VAI · 3 NAI
Inspections with 483 observations shown
4 of 7
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Nov 1, 2024
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Equipment control activity documentation

      Equipment calibrations have not been documented.

    2. Records, DHR {see also 820.184(e)}

      Labels and labeling used for each finished product, lot, or batch, were not sufficiently documented in the DHR.

    3. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      In-process inspections, tests, or other verification activities and approvals were not documented.

    2. Lack of procedures, or not maintained

      Document control procedures have not been maintained.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    2. Retention period

      Required records are not retained for the design and expected life of the device.

    3. Supplier oversight

      The type and extent of control to be exercised over suppliers was not clearly defined.

    4. Lack of or inadequate procedures for distribution

      Procedures for control and distribution of finished devices have not been adequately established.

    5. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    Show 4 more citations
    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    2. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    3. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    4. Equipment Installation, Placement, Specified Requirements

      The appropriate design, construction, placement, and installation of manufacturing equipment have not been ensured.

  6. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  7. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - simulated testing

      The design was not validated using production units under actual or simulated use conditions.

    2. Design control - no procedures

      Procedures to control the design process of the device were not established, complete, and implemented.

    3. Management review - Lack of or inadequate procedures

      Procedures for management review are not implemented.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Nov 1, 2024. Each one asked for a Form 483.

  • Inspection ending Sep 19, 2024VAI

    2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Nov 1, 2024FDA-FOIA-2024-9653Form 483Closed
    2. Oct 25, 2024FDA-FOIA-2024-9514Form 483· 1 of 2 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 13 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 7 inspections, with how many of them cited each.

Show all 13

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1221765 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1221765. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.