Optim LLC
64 Technology Park Rd, Sturbridge, Massachusetts, United States
FEI 1218141
FDA has 4 inspections on file for Optim LLC in Sturbridge, Massachusetts (FEI 1218141), from 2010 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.198(a), 21 CFR 820.100(a) and 21 CFR 820.40.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2010–2025
- 0 OAI · 4 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Jan 27, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design history file
The design history file was not established and does not demonstrate that the design was developed following the approved design plan.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures not adequately established or maintained
Document control procedures have not been adequately maintained.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely inspected, checked, and maintained have not been adequately established.
Documentation of software validation
Software validation activities and results for computers or automated data processing systems used as part of production and the quality system have not been documented.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 5 more citations
Sampling plans
Sampling plans are not based on valid statistical rationale.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Procedures not adequately established or maintained
Document control procedures have not been adequately maintained. Electronic records are used, but they do not meet system access limitation requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Review and evaluation for investigation
Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Jan 27, 2026. Each one asked for a Form 483.
Inspection ending May 29, 2025VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jan 30, 2026. FDAgent holds no copy that FDA posted.
- Oct 14, 2025FDA-FOIA-2025-11852Form 483· 1 of 5 sitesRecords released
Inspection ending Oct 28, 2021VAI
1 request. In the inspection history above.
- Jan 27, 2026FDA-FOIA-2026-795Form 483· 1 of 5 sitesOpen
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls3 of 4
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 820.40[Reserved]2 of 4
- 21 CFR 820.100Corrective and preventive action2 of 4
- 21 CFR 820.20Management responsibility1 of 4
- 21 CFR 820.22Quality audit1 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.70Production and process controls1 of 4
Show all 14
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
- 21 CFR 820.184Device history record1 of 4
- 21 CFR 820.200Servicing1 of 4
- 21 CFR 820.250Statistical techniques1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1218141 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1218141. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.