Skip to content

CooperSurgical, Inc.

95 Corporate Dr, Trumbull, Connecticut, United States

FEI 1216677DUNS 118670081

Company: CooperSurgical, Inc. · Linked by exact D-U-N-S match (801895244) in FDA registration data

FDA has 15 inspections on file for CooperSurgical, Inc. in Trumbull, Connecticut (FEI 1216677), from 2009 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published one citation from this inspection, under 21 CFR Part 211, the manufacturing rules for finished drugs. It is listed below. One FOIA request has asked FDA for this inspection's 483.

The most recent OAI inspection ended on Dec 17, 2014. FDA has issued 2 warning letters under this FEI, the latest on (CDER · The Office of Prescription Drug Promotion (OPDP)). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 803.50(a)(2), 21 CFR 820.198(a) and 21 CFR 820.30(g).

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, packing, analysis, relabeling and repacking, through Dec 31, 2026. Its registrant, CooperSurgical, Inc. (D-U-N-S 801895244), has registered 1 other site.

Sources:FDA firm profile ↗Warning letter ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
15, 2009–2024
1 OAI · 6 VAI · 8 NAI
Inspections with 483 observations shown
7 of 15
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Jul 3, 2024
Import refusals
1 refusal
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

15 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Storage under appropriate conditions

      Drug products are not stored under appropriate conditions of temperature and humidity so that their identity, strength, quality, and purity are not affected.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    2. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    3. Procedures for product review,disposition lack of/inadequate

      Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.

    4. Mixups

      Labeling and packaging operations were not controlled to prevent labeling mix-ups.

  4. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  5. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not fully followed.

  6. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  7. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    3. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    4. Sampling methods - Lack of or inadequate procedures

      Procedures to ensure sampling methods are adequate for their intended use have not been adequately established.

    5. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    Show 3 more citations
    1. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Design validation - simulated testing

      The design was not validated under actual or simulated use conditions.

  8. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  9. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  10. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Complaint Investigation/Follow-Up Findings

      Complaint records are deficient in that they do not include the findings of the investigation and follow-up.

  11. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Project Evaluation: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - simulated testing

      The design was not validated under actual or simulated use conditions.

    2. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  12. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  13. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  14. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  15. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Expiration date lacking

      Drug products do not bear an expiration date determined by appropriate stability data to assure they meet applicable standards of identity, strength, quality and purity at the time of use.

    2. Yield calculations not verified by 2nd person

      Yield calculations are not performed by one person and independently verified by a second person.

    3. Actual yield, % of theoretical yield

      The batch production and control records do not include a statement of the actual yield and percentage of theoretical yield at appropriate stages of processing for each batch of drug product produced.

    4. Lack of written stability program

      There is no written testing program designed to assess the stability characteristics of drug products.

    5. Test methods

      The accuracy, sensitivity, specificity, and reproducibility of test methods have not been established and documented.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Jul 3, 2024. It asked for a Form 483.

  • Inspection ending May 21, 2024VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Jul 3, 2024FDA-FOIA-2024-5867Form 483· 1 of 2 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 17 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 15 inspections, with how many of them cited each.

Show all 17

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1216677, one row per refused product line.

1 refusal on 2019-03-15 · charged NOT LISTED on 1, FRNMFGREG on 1 · no FDA sample analysis

FDA firm profile
  • 2019-03-15

    REMOVER, INTRAUTERINE DEVICE, CONTRACEPTIVE, HOOK-TYPE85HHF

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-0380053-8/11/1

FDA drug establishment registration

FDA registration search
Registered operations
Manufacturing, packing, analysis, relabeling, repacking
Registration runs through
Dec 31, 2026
Registrant
CooperSurgical, Inc.D-U-N-S 801895244

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Other sites with the same registrant

FDA’s registration file names CooperSurgical, Inc. (D-U-N-S 801895244) as the registrant for this site and the one other below. That exact D-U-N-S number is the only link. A registrant is the firm that filed the registration, which is not always the owner.

  • CooperSurgical, Inc.North Tonawanda, New York · FEI 13166269 inspections9latestFeb 1, 2024VAINone on file

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. This site’s records are filed under FEI 1216677 in FDA’s data; other sites under its registrant are listed in summary only. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every inspection and enforcement record on this page is filed under FEI 1216677. Other sites filed under the same registrant D-U-N-S are listed in summary only. Records filed under another FEI or name are otherwise not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.