Sea-Long Medical Systems, LLC
2498 Fm 66 Building 1021, Waxahachie, Texas, United States
FEI 1037230
FDA has 3 inspections on file for Sea-Long Medical Systems, LLC in Waxahachie, Texas (FEI 1037230), from 2014 to 2024. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 11 citations from this inspection: 10 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 801, the device labeling rules. They are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.
FDA issued a warning letter on (CDRH · Division of Medical Device and Radiological Health Operations West). No court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 3, 2014–2024
- 1 OAI · 0 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Jun 25, 2025
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
DHR - not or inadequately maintained
A device history record has not been maintained.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been established.
Show 6 more citations
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DMR - not or inadequately maintained
A device master record has not been maintained.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Show 6 more citations
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Jun 25, 2025. Each one asked for a Form 483.
Inspection ending Nov 5, 2024OAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Jun 25, 2025FDA-FOIA-2025-8004Form 483· 1 of 2 sitesClosed over fees
- Dec 20, 2024FDA-FOIA-2024-11207Form 483· 1 of 2 sitesClosed
Inspection ending Sep 8, 2022NAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- May 16, 2024FDA-FOIA-2024-4451Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit2 of 2
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.75Process validation2 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.184Device history record2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 801.20Label to bear a unique device identifier1 of 2
Show all 13
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.25Personnel1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.181Device master record1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · Division of Medical Device and Radiological Health Operations West · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1037230 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1037230. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.