Skip to content

Linvatec Corp. dba ConMed Linvatec

11311 Concept Blvd, Largo, Florida, United States

FEI 1017294

FDA has 6 inspections on file for Linvatec Corp. dba ConMed Linvatec in Largo, Florida (FEI 1017294), from 2009 to 2023. The latest ended on and was classified NAI, no action indicated. FDA has also posted a Form 483 dated Mar 24, 2026 that FDAgent has not matched to an inspection on the Data Dashboard.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

One FOIA request asked FDA for the 483 from a later inspection, ending Mar 24, 2026, that FDA's Data Dashboard doesn't list yet.

The most recent OAI inspection ended on Mar 26, 2009. 6 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
6, 2009–2023
1 OAI · 2 VAI · 3 NAI
Inspections with 483 observations shown
3 of 6
1 posted 483 not matched to an inspection
Form 483s posted by FDA
↗
Requested under FOIA
1 request, latest May 27, 2026
Warning letters
None on file
Import refusals
6 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC

  1. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of System for Determining MDR Events

      The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and implemented.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    5. Records, DHR {see also 820.184(e)}

      Labels and labeling used for each finished product, lot, or batch, were not sufficiently documented in the DHR.

    Show 2 more citations
    1. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers, contractors, and consultants have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design input procedures - appropriateness

      Procedures to ensure that a device's design input requirements are appropriate and address its intended use, including user/patient needs, were not implemented.

    2. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    3. Documentation - review in response to changes or deviations

      There is no documentation of the review and evaluation of a process and revalidation of a process conducted in response to changes or process deviations.

  7. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate In-process acceptance procedures

      Acceptance procedures to ensure that specified requirements for in-process product are met were not complete.

    2. Investigation

      Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.

    3. Documentation of disposition, justification, signature

      There is incomplete documentation of the disposition of nonconforming product.

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on May 27, 2026. It asked for a Form 483.

  • Inspection ending Mar 24, 2026

    1 request. Not listed in FDA’s Data Dashboard yet. FDA has posted its Form 483.

    1. May 27, 2026FDA-FOIA-2026-5300Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 6 inspections, with how many of them cited each.

Show all 10

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1017294, one row per refused product line.

6 refusals from 2018-03-14 to 2025-09-05 · charged FRNMFGREG on 5, NOT LISTED on 2 · no FDA sample analysis

FDA firm profile
  • 2025-09-05

    DRILL, BONE, POWERED76DZI

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsAEK-7471781-8/10001/1

  • 2019-05-31

    SAW, POWERED, AND ACCESSORIES79HAB

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-1644611-3/21/1

  • 2019-05-31

    SAW, POWERED, AND ACCESSORIES79HAB

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-1644611-3/11/1

  • 2019-01-30

    ANCHOR, FASCIAL79NEH

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-9505380-5/11/1

  • 2018-03-14

    SAW, POWERED, AND ACCESSORIES79HAB

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-4962036-0/11/2

Show 1 more
  • 2018-03-14

    DRILL, BONE, POWERED76DZI

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-4962036-0/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1017294 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1017294. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.