Magnificent Medical Equipment, Inc. DBA Tony Riso Company

10015 Se Osprey Pointe Dr, Hobe Sound, Florida, United States

FEI 1000526376

FDA has one inspection on file for Magnificent Medical Equipment, Inc. DBA Tony Riso Company in Hobe Sound, Florida (FEI 1000526376). It ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 14 citations from this inspection: 13 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2009
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices14 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Production processes

      Production processes were not complete to ensure that a device conforms to its specifications.

    2. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not established.

    3. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated and approved according to established procedures.

    5. General

      Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been established and documented.

    Show 9 more citations
    1. Evaluation of suppliers, contractors, etc., requirements

      Adequate quality requirements that must be met by contractors were not established.

    2. Design control - no procedures

      Procedures to control the design process of the device were not established.

    3. Acceptance records

      The device history record does not include acceptance records that demonstrate the device is manufactured in accordance with the device master record.

    4. Incoming inspection, testing, verification

      Incoming product was not adequately inspected or tested to verify conformance to specifications.

    5. Results of investigation

      Records of complaint investigations do not include the results of the investigation.

    6. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure that equipment is routinely calibrated were not established, documented, and implemented.

    7. Documentation

      Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities, the actions needed to correct or prevent recurrence of nonconforming product and other quality problems, the verification or validation of corrective actions, and implementation of corrective and preventive actions.

    8. Lack of procedures, or not maintained

      Document control procedures were not established.

    9. DMR QA procedures and specifications

      The device master record does not include or refer to the location of quality assurance procedures and specifications.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Show all 12

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000526376 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000526376. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.