Circulatory Technology Inc.
21 Singworth St, Oyster Bay, New York, United States
FEI 1000522036
FDA has 7 inspections on file for Circulatory Technology Inc. in Oyster Bay, New York (FEI 1000522036), from 2010 to 2019. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 7 citations from this inspection: 6 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
The most recent OAI inspection ended on Sep 26, 2018. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.
10 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 7, 2010–2019
- 2 OAI · 3 VAI · 2 NAI
- Inspections with 483 observations shown
- 5 of 7
- Form 483s posted by FDA
- None held for this FEI
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of procedures, or not maintained
Document control procedures have not been established.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers and contractors were not evaluated based on their ability to meet specified requirements.
Design validation - Risk analysis
Results of the design risk analysis were not adequately documented.
Show 2 more citations
Design history file
The design history file was not established.
Process changes - review, evaluation and revalidation
A validated process was not revalidated when changes or process deviations occurred.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Acceptance status
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria.
Design validation - documentation
The results of design validation, including identification of the design, method(s), the date and the individual(s) performing validation, were not documented in the design history file.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Design history file
The design history file does not demonstrate that the design was developed following the approved design plan and does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Show 4 more citations
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.
Process changes - review, evaluation and revalidation
A validated process was not revalidated when changes or process deviations occurred.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated and investigated where necessary.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers and contractors were not evaluated based on their ability to meet specified requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documented evaluation
The evaluation of potential suppliers and contractors was not documented.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Design validation acceptance criteria
Acceptance criteria were not established prior to the performance of validation activities.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
MDR not evaluated per 820.198
A submitted MDR event was not evaluated in accordance with the requirements of the regulations regarding Failure Investigations and Complaint Files.
Show 5 more citations
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.
Design review - documentation
The design review results, including identification of the design, the date and the individual(s) performing the review , were not documented in the design history file.
Acceptance status
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria.
Design history file
The design history file was not established and does not demonstrate that the design was developed following the requirements of 21 CFR 820.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Acceptable supplier records, inadequate records
Records of acceptable suppliers, contractors, and consultants have not been adequately established.
Design control - no procedures
Procedures for design control have not been established.
Design review - documentation
The design review results, including identification of the design, the date, and the individual(s) performing the review, were not documented in the design history file.
Design validation acceptance criteria
Acceptance criteria were not established prior to the performance of validation activities.
Show 1 more citation
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
General
Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been established, defined, documented, completed, and implemented.
Time to report - 10 days
A report of the required information regarding device correction and removal actions was not sent to FDA within 10 days of initiating the correction or removal.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential contractors and consultants were not sufficiently evaluated and selected on the basis of their ability to meet specified requirements.
Design changes - Lack of or Inadequate Procedures
Procedures were not established, defined, documented, completed, and followed for the identification, documentation, validation or verification, review, and approval of design changes before their implementation.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained, and implemented.
Show 1 more citation
Design validation acceptance criteria
Acceptance criteria were not established, defined, documented, complete, and implemented prior to the performance of validation activities.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls5 of 5
- 21 CFR 820.50Purchasing controls5 of 5
- 21 CFR 820.100Corrective and preventive action4 of 5
- 21 CFR 820.198Complaint files4 of 5
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?3 of 5
- 21 CFR 820.75Process validation2 of 5
- 21 CFR 820.86Acceptance status2 of 5
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 5
Show all 11
- 21 CFR 803.18What are the requirements for establishing and maintaining MDR files or records that apply to me?1 of 5
- 21 CFR 806.10Reports of corrections and removals1 of 5
- 21 CFR 820.40[Reserved]1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000522036 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000522036. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.