OcuMetrics, Inc.
3301 Middlefield Rd, Palo Alto, California, United States
FEI 1000221346
FDA has 4 inspections on file for OcuMetrics, Inc. in Palo Alto, California (FEI 1000221346), from 2009 to 2019. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 820.50(a)(2) and 21 CFR 820.22.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2009–2019
- 0 OAI · 1 VAI · 3 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Supplier oversight
The type and extent of control to be exercised over suppliers and consultants was not clearly defined.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a procedure have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Supplier oversight
The type and extent of control to be exercised over services, suppliers, and consultants was not clearly defined.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been established.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Acceptable supplier records, inadequate records
Records of acceptable suppliers and contractors have not been established.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
Show 2 more citations
Installer records
The person installing the device did not adequately document the inspection and test results to demonstrate proper installation.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - software validation not performed
Validation of device software is incomplete.
Design validation- Lack of or inadequate procedures
Procedures for validating the device design were not complete.
Documentation
Acceptance test results of in-process product were not fully documented.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls3 of 4
- 21 CFR 820.50Purchasing controls3 of 4
- 21 CFR 820.22Quality audit2 of 4
- 21 CFR 820.100Corrective and preventive action2 of 4
- 21 CFR 820.70Production and process controls1 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
- 21 CFR 820.170Installation1 of 4
Show all 11
- 21 CFR 820.184Device history record1 of 4
- 21 CFR 820.198Complaint files1 of 4
- 21 CFR 820.200Servicing1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000221346 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000221346. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.