North Coast Medical Inc
135 E Main Ave Ste 110, Morgan Hill, California, United States
FEI 1000124207
FDA has 6 inspections on file for North Coast Medical Inc in Morgan Hill, California (FEI 1000124207), from 2009 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on May 25, 2011. 1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2009–2024
- 1 OAI · 3 VAI · 2 NAI
- Inspections with 483 observations shown
- 4 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Nov 3, 2023
- Warning letters
- None on file
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Review and evaluation for investigation
Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Training records
Personnel training is not documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Design validation - simulated testing
The design was not validated under actual or simulated use conditions.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Show 2 more citations
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Retention period
Required records are not retained for at least 2 years from the date of release of the device for commercial distribution.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Show 5 more citations
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Design validation - simulated testing
The design was not validated under actual or simulated use conditions.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Nov 3, 2023. It asked for a Form 483.
Inspection ending Aug 28, 2023VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Nov 3, 2023FDA-FOIA-2023-9757Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.25Personnel3 of 4
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 806.10Reports of corrections and removals2 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.40[Reserved]2 of 4
- 21 CFR 820.50Purchasing controls2 of 4
- 21 CFR 820.90Nonconforming product2 of 4
- 21 CFR 820.100Corrective and preventive action2 of 4
Show all 12
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 4
- 21 CFR 820.22Quality audit1 of 4
- 21 CFR 820.180General requirements1 of 4
- 21 CFR 820.184Device history record1 of 4
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1000124207, one row per refused product line.
1 refusal on 2023-12-08 · charged NOT LISTED on 1 · no FDA sample analysis
FDA firm profile- 2023-12-08
SPLINT, HAND, AND COMPONENTS89ILH
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of West Coast Imports799-8099457-8/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000124207 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000124207. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.