Skip to content

Koko, LLC

1200 S Fordham St Ste B, Longmont, Colorado, United States

FEI 1000117266

FDA has 4 inspections on file for Koko, LLC in Longmont, Colorado (FEI 1000117266), from 2010 to 2021. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 9 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.198(a) and 21 CFR 820.30(g).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2010–2021
0 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
3 of 4
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
1 refusal
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Software validation for automated processes

      Software used as part of production has not been validated for its intended use according to an established protocol.

    2. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    5. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a specification, method, process and procedure have not been adequately established.

    Show 4 more citations
    1. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    3. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate instructions

      Installation instructions have not been established.

    2. Training records

      Personnel training is not documented.

    3. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    4. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    5. Remedial action - documentation

      Evaluations of out-of-calibration equipment and remedial actions taken were not documented.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures for product review,disposition lack of/inadequate

      Procedures that describe the review and disposition process for nonconforming product were not defined and implemented.

    2. DHR content

      The device history record does not demonstrate the device is manufactured in accordance with the device master record.

    3. General

      Complaint handling procedures for receiving and evaluating complaints have not been completed.

    4. Analysis of data sources

      Appropriate sources of quality data are not adequately analyzed to identify existing and potential causes of nonconforming product and other quality problems.

    5. Nonconforming product, Lack of or inadequate procedures

      Procedures were not followed for the control of products that do not conform to specifications.

    Show 1 more citation
    1. Lack of or inadequate DHR procedures

      Procedures have not been completed and implemented to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.

Show all 12

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1000117266, one row per refused product line.

1 refusal on 2020-01-14 · charged NO 510(K) on 1 · no FDA sample analysis

FDA firm profile
  • 2020-01-14

    SPIROMETER, DIAGNOSTIC73BZG

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of West Coast Imports799-1414660-8/21/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000117266 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000117266. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.