Skip to content

21 CFR 1010.3: Identification

FDA's inspection data lists 26 Form 483 citations of 21 CFR 1010.3 at 22 facilities, from May 2010 to June 6, 2025.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
26
At 23 inspections
Facilities cited
22
Distinct FEIs
Most recent citation
Jun 6, 2025
Transdermal Cap Inc.
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 1010.3 from the Electronic Code of Federal Regulations.

§ 1010.3 Identification.

(a) Every manufacturer of an electronic product to which a standard under this subchapter is applicable shall set forth the information specified in paragraphs (a)(1) and (2) of this section. This information shall be provided in the form of a tag or label permanently affixed or inscribed on such product so as to be legible and readily accessible to view when the product is fully assembled for use or in such other manner as may be prescribed in the applicable standard. Except for foreign equivalent abbreviations as authorized in paragraph (a)(1) of this section all such labels or tags shall be in the English language.

(1) The full name and address of the manufacturer of the product; abbreviations such as “Co.,” “Inc.,” or their foreign equivalents and the first and middle initials of individuals may be used. Where products are sold under a name other than that of the manufacturer of the product, the full name and address of the individual or company under whose name the product was sold may be set forth, provided such individual or company has previously suppled the Director, Center for Devices and Radiological Health with sufficient information to identify the manufacturer of the product.

(2) The place and month and year of manufacture:

(i) The place of manufacture may be expressed in code provided the manufacturer has previously supplied the Director, Center for Devices and Radiological Health with the key to such code.

(ii) The month and year of manufacture shall be provided clearly and legibly, without abbreviation, and with the year shown as a four-digit number as follows in this paragraph. Alternatively, a manufacturer may utilize a manufacturing symbol and date format that conforms with an applicable FDA recognized consensus standard.

Manufactured: (Insert Month and Year of Manufacture.)

(b) In the case of products for which it is not feasible to affix identification labeling in accordance with paragraph (a) of this section, upon application by the manufacturer, the Director, Center for Devices and Radiological Health may approve an alternate means by which such identification may be provided.

(c) Every manufacturer of an electronic product to which a standard under this subchapter is applicable shall provide to the Director, Center for Devices and Radiological Health a list identifying each brand name which is applied to the product together with the full name and address of the individual or company for whom each product so branded is manufactured.

21 CFR Part 1010: Performance Standards for Electronic Products: General, Subpart ARead on eCFR

Citations by year

Form 483 citations of 21 CFR 1010.3 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 4 in 2011

Form 483 citations of 21 CFR 1010.3 by inspection year
YearForm 483 citations
20102
20114
20131
20153
20161
20171
20181
20194
20201
20224
20242
20252

How FDA words this citation

FDA's short descriptions and citation text on these 26 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • ID label lacks name and address8 citations
  • ID label lacks place, month & year of manufacture8 citations
  • ID label or tag5 citations
  • Manufacture date abbreviated or unapproved date coding5 citations

Citation text

  • “An identification label failed to provide the name and address of the manufacturer.”5 citations
  • “An identification label fails to provide the month and year of manufacture.”4 citations
  • “An identification label fails to provide the place, month and year of manufacture.”3 citations
  • “An identification label or tag is not permanently affixed or inscribed.”3 citations
Show 2 more
  • “The date of manufacture provided on the identification label is not stated in the prescribed manner.”3 citations
  • “An identification label or tag is not permanently affixed or inscribed and readily accessible to view when the product is fully assembled for use.”2 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 1010.3, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

Show 4 more
  • Ultimate F/X Lasers and Pyrotechnics LLC2022-05-05 · FEI 3021402469ID label lacks place, month & year of manufacture
  • Inova Lasers LLC2020-02-07 · FEI 3015191149ID label lacks place, month & year of manufacture
  • KBD, Inc.2019-11-08 · FEI 1022129Manufacture date abbreviated or unapproved date coding
  • CureWave Lasers, LLC2019-09-05 · FEI 3014268427ID label lacks name and address

Often cited at the same inspection

Of the 23 inspections where FDA cited 21 CFR 1010.3, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 1010.3

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.