Epicare Acquisitions, Llc
11 Princess Rd Ste C, Lawrenceville, New Jersey, United States
FEI 3015730775
FDA has 2 inspections on file for Epicare Acquisitions, Llc in Lawrenceville, New Jersey (FEI 3015730775), from 2022 to 2024. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 10 citations from this inspection. They are listed below. Three FOIA requests have asked FDA for this inspection's 483, and one more for the 483 from a later inspection, ending Jan 12, 2026, that FDA's Data Dashboard doesn't list yet.
No court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.184 and 21 CFR 820.25(b).
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 2, 2022–2024
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Mar 18, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety Act10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 5 more citations
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Manufacture date abbreviated or unapproved date coding
The date of manufacture provided on the identification label is abbreviated.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
Design verification - documentation
The design verification results, including method(s) and the date, were not adequately documented in the design history file.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Mar 18, 2026. Each one asked for a Form 483.
Inspection ending Jan 12, 2026
1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Aug 27, 2026. FDAgent holds no copy that FDA posted.
- Mar 18, 2026FDA-FOIA-2026-2720Form 483· 1 of 5 sitesRecords released
Inspection ending Dec 13, 2024OAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 31, 2025. FDAgent holds no copy that FDA posted.
- Sep 29, 2025FDA-FOIA-2025-11412Form 483· 1 of 4 sitesRecords released
- Jul 31, 2025FDA-FOIA-2025-9265Form 483Records released
- Apr 30, 2025FDA-FOIA-2025-3888Form 483· 1 of 4 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.25Personnel2 of 2
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.184Device history record2 of 2
- 21 CFR 801.20Label to bear a unique device identifier1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.181Device master record1 of 2
- 21 CFR 820.198Complaint files1 of 2
Show all 10
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 2
- 21 CFR 1010.3Identification1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3015730775 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3015730775. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.