Skip to content

Inova Lasers LLC

433 5th Ave Fl 6, New York, New York, United States

FEI 3015191149

FDA has 2 inspections on file for Inova Lasers LLC in New York, New York (FEI 3015191149), from 2020 to 2021. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection. They are listed below.

The most recent OAI inspection ended on Feb 7, 2020. No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 1010.2(b).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
2, 2020–2021
1 OAI · 1 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Aperture label

      A label bearing the wording: "Laser aperture". is not affixed to the medical laser product, in close proximity to the aperture through which was emitted accessible laser radiation in excess of the accessible emission limits of Class I.

    2. Certification other than by tag or label

      Certification of a product is not in conformance with the alternate means approved by FDA.

    3. Legible reproductions

      Legible reproductions of all labels and hazard warnings required by the regulations were not provided with the laser product.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety Act11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Failure to submit, distinct marking

      You did not submit a required product report prior to the introduction of the product into commerce.

    2. Key control

      Each laser system classified as a Class IIIb or Class IV] laser product did not incorporate a[key-actuated master control

    3. Certification label or tag

      A certification label or tag is not permanently affixed or inscribed.

    4. Incoming acceptance records, documentation

      Acceptance or rejection of incoming product was not documented.

    5. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    Show 6 more citations
    1. Calibration of measurement system

      The Class IV medical laser product is not supplied with instructions specifying a procedure and schedule for calibration of the measurement system.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    3. ID label lacks place, month & year of manufacture

      An identification label fails to provide the place of manufacture.

    4. Certification not based on adequate test/testing program

      Certification was not based upon a testing program in accordance with good manufacturing practices.

    5. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

    6. Class IV "Danger" label

      Each Class IV laser product does not have affixed a label bearing the DANGER logotype specified in the regulation and bearing the wording "CLASS IV LASER PRODUCT" in the position specified in the regulation.

Regulations FDA cited here

The 9 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Show all 9

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3015191149 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3015191149. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.