21 CFR 1040.11: Specific purpose laser products

FDA's inspection data lists 24 Form 483 citations of 21 CFR 1040.11 at 18 facilities, from June 2010 to May 2, 2025.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
24
At 20 inspections
Facilities cited
18
Distinct FEIs
Most recent citation
May 2, 2025
Focus Medical, LLC
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 1040.11 from the Electronic Code of Federal Regulations.

§ 1040.11 Specific purpose laser products.

(a) Medical laser products. Each medical laser product shall comply with all of the applicable requirements of § 1040.10 for laser products of its class. In addition, the manufacturer shall:

(1) Incorporate in each Class III or IV medical laser product a means for the measurement of the level of that laser radiation intended for irradiation of the human body. Such means may have an error in measurement of no more than 20 percent when calibrated in accordance with paragraph (a)(2) of this section. Indication of the measurement shall be in International System Units. The requirements of this paragraph do not apply to any laser radiation that is all of the following:

(i) Of a level less than the accessible limits of Class IIIa; and

(ii) Used for relative positioning of the human body; and

(iii) Not used for irradiation of the human eye for ophthalmic purposes.

(2) Supply with each Class III or IV medical laser product instructions specifying a procedure and schedule for calibration of the measurement system required by paragraph (a)(1) of this section.

(3) Affix to each medical laser product, in close proximity to each aperture through which is emitted accessible laser radiation in excess of the accessible emission limits of Class I, a label bearing the wording: “Laser aperture.”

(b) Surveying, leveling, and alignment laser products. Each surveying, leveling. or alignment laser product shall comply with all of the applicable requirements of § 1040.10 for a Class I, IIa, II or IIIa laser product and shall not permit human access to laser radiation in excess of the accessible emission limits of Class IIIa.

(c) Demonstration laser products. Each demonstration laser product shall comply with all of the applicable requirements of § 1040.10 for a Class I, IIa, II, or IIIa laser product and shall not permit human access to laser radiation in excess of the accessible emission limits of Class I and, if applicable, Class IIa, Class II, or Class IIIa.

21 CFR Part 1040: Performance Standards for Light-Emitting ProductsRead on eCFR

Citations by year

Form 483 citations of 21 CFR 1040.11 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 4 in 2018

Form 483 citations of 21 CFR 1040.11 by inspection year
YearForm 483 citations
20101
20111
20121
20132
20151
20161
20184
20193
20202
20214
20222
20252

How FDA words this citation

FDA's short descriptions and citation text on these 24 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Aperture label8 citations
  • Demonstration laser products6 citations
  • Calibration of measurement system4 citations
  • Med. laser compliance, measurement4 citations
  • Surveying, leveling, alignment laser products2 citations

Citation text

  • “A label bearing the wording: "Laser aperture". is not affixed to the medical laser product, in close proximity to the aperture through which was emitted accessible laser radiation in excess of the accessible emission limits of Class I.”8 citations
  • “Each demonstration laser product does not comply with all of the applicable requirements for a Class I, IIa, II, or IIIa laser product and prevent human access to laser radiation in excess of the accessible emission limits of Class I, IIa, II or IIIa.”3 citations
  • “The Class IV medical laser product is not supplied with instructions specifying a procedure and schedule for calibration of the measurement system.”3 citations
  • “The medical laser product does not incorporate in each Class III or IV medical laser product a means for the measurement of the level of that laser radiation intended for irradiation of the human body, having an error in measurement of no more than 20% when calibrated in accordance with the CFR requirements.”3 citations
Show 2 more
  • “Each demonstration laser product does not comply with all of the applicable requirements for a Class I, IIa, II, or IIIa laser product.”2 citations
  • “Each demonstration laser product does not prevent human access to laser radiation in excess of the accessible emission limits of Class I, IIa, II or IIIa.”1 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 1040.11, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

Show 4 more

Often cited at the same inspection

Of the 20 inspections where FDA cited 21 CFR 1040.11, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 1040.11

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 1040.11; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.