WEBER MEDICAL GMBH
Sohnreystr. 4, Lauenforde, Germany
FEI 3007165178
FDA has 3 inspections on file for WEBER MEDICAL GMBH in Lauenforde, Germany (FEI 3007165178), from 2013 to 2022. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued. One FOIA request has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Nov 8, 2018. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2013–2022
- 2 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 1 of 3
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Aug 6, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety Act15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Legible reproductions
Legible reproductions of all labels and hazard warnings required by the regulations were not provided with the laser product and did not include the information required for positions 1, 2, and 3 of the logotype specified in the CFR.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Emission indicator - Class IIIb & IV
An emission indicator was not incorporated in each laser system classified as a Class IIIb or Class IV laser product, which provides a visible or audible signal during emission of accessible laser radiation in excess of the accessible emission limits of Class I and which signals sufficiently prior to emission of such radiation, to allow appropriate action to avoid exposure to the laser radiation.
Aperture label
A label bearing the wording: "Laser aperture". is not affixed to the medical laser product, in close proximity to the aperture through which was emitted accessible laser radiation in excess of the accessible emission limits of Class I.
Failure to submit
You did not submit an annual report for products requiring one.
Show 10 more citations
Reproduction of affixed information
The manufacturer did not provide or cause to be provided in all catalogs pertaining to each laser product, a legible reproduction of the class designation, the warning and the information required for positions 1, 2, and 3 of the applicable logotype required b to be affixed to the product.
Certification label or tag
A certification label or tag is not permanently affixed or inscribed.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of System for Timely Submission of Reports
The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Justification for not reporting
A justification for not reporting the correction or removal action to FDA that included conclusions and reviews by a designated person was not included in the record.
User info - list controls, etc.
The manufacturer did not provide or cause to be provided a listing of the warning: "Caution --- use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.".
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Aug 6, 2026. It asked for a Form 483.
Inspection ending Sep 22, 2022VAI
1 request. In the inspection history above.
- Aug 6, 2026FDA-FOIA-2026-7585Form 483· 1 of 2 sitesOpen
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 3 inspections.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
- 21 CFR 820.90Nonconforming product1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.198Complaint files1 of 1
Show all 12
- 21 CFR 1002.13Annual reports1 of 1
- 21 CFR 1010.2Certification1 of 1
- 21 CFR 1040.10Laser products1 of 1
- 21 CFR 1040.11Specific purpose laser products1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3007165178 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3007165178. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.