CureWave Lasers, LLC
11915 Quincy Ln, Dallas, Texas, United States
FEI 3014268427
FDA has 2 inspections on file for CureWave Lasers, LLC in Dallas, Texas (FEI 3014268427), from 2019 to 2021. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 7 citations from this inspection. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 2, 2019–2021
- 0 OAI · 0 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest FY2023
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Certification not based on adequate test/testing program
Certification was not based upon a test, in accordance with the standard and a testing program in accordance with good manufacturing practices.
Supplier oversight
The type and extent of control to be exercised over the product, services, suppliers, contractors and consultants was not clearly defined.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected and maintained have not been established.
Aperture label
A label bearing the wording: "Laser aperture". is not affixed to the medical laser product, in close proximity to the aperture through which was emitted accessible laser radiation in excess of the accessible emission limits of Class I.
Show 2 more citations
Calibration of measurement system
The Class IV medical laser product is not supplied with instructions specifying a procedure and schedule for calibration of the measurement system.
Med. laser compliance, measurement
The medical laser product does not incorporate in each Class III or IV medical laser product a means for the measurement of the level of that laser radiation intended for irradiation of the human body, having an error in measurement of no more than 20% when calibrated in accordance with the CFR requirements.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers, contractors and consultants were not evaluated based on their ability to meet specified requirements.
Documentation
Acceptance activities were not documented and maintained as part of the device history record.
ID label lacks name and address
An identification label failed to provide the name and address of the individual or company under whose name the product was sold.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Lack of procedures, or not maintained
Document control procedures have not been established.
Show 7 more citations
Certification not based on adequate test/testing program
Certification was not based upon a test, in accordance with the standard.
Med. laser compliance, measurement
The medical laser product does not incorporate in each Class III or IV medical laser product a means for the measurement of the level of that laser radiation intended for irradiation of the human body, having an error in measurement of no more than 20% when calibrated in accordance with the CFR requirements.
Aperture label
A label bearing the wording: "Laser aperture". is not affixed to the medical laser product, in close proximity to the aperture through which was emitted accessible laser radiation in excess of the accessible emission limits of Class I.
DHR - not or inadequately maintained
A device history record has not been maintained.
Design history file
The design history file was not established.
Failure to submit
You did not submit an annual report by the September 1 deadline for products requiring one.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.
Inspection ending Aug 5, 2021NAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-5194Form 483Closed
Inspection ending Sep 5, 2019NAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 25, 2026. FDAgent holds no copy that FDA posted.
- Aug 3, 2026FDA-FOIA-2026-7393Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.40[Reserved]2 of 2
- 21 CFR 820.50Purchasing controls2 of 2
- 21 CFR 1010.2Certification2 of 2
- 21 CFR 1040.11Specific purpose laser products2 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
Show all 12
- 21 CFR 820.181Device master record1 of 2
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 1002.13Annual reports1 of 2
- 21 CFR 1010.3Identification1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3014268427 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3014268427. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.