Focus Medical, LLC
23 Francis J Clarke Cir, Bethel, Connecticut, United States
FEI 3003067608
FDA has 6 inspections on file for Focus Medical, LLC in Bethel, Connecticut (FEI 3003067608), from 2010 to 2025. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 8 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.198(a), 21 CFR 820.184 and 21 CFR 1002.13.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 6, 2010–2025
- 1 OAI · 3 VAI · 2 NAI
- Inspections with 483 observations shown
- 4 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Jan 9, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety Act8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of to meet any of its specifications were not investigated where necessary.
Aperture label
A label bearing the wording: "Laser aperture". is not affixed to the medical laser product, in close proximity to the aperture through which was emitted accessible laser radiation in excess of the accessible emission limits of Class I.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.
Class IV "Danger" label
Each Class IV laser product does not have affixed a label bearing the DANGER logotype specified in the regulation, bearing the wording "LASER RADIATION--AVOID EYE OR SKIN EXPOSURE TO DIRECT OR SCATTERED RADIATION" in the position specified in the regulation and bearing the wording "CLASS IV LASER PRODUCT" in the position specified in the regulation.
Show 3 more citations
ID label lacks name and address
An identification label failed to provide the name and address of the.
Legible reproductions
Legible reproductions of all labels and hazard warnings required by the regulations were not provided with the laser product.
Calibration of measurement system
The Class IV medical laser product is not supplied with instructions specifying a procedure and schedule for calibration of the measurement system.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Failure to submit
You did not submit an annual report for products requiring one.
Personnel
Personnel do not have the necessary training to perform their jobs.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documented evaluation
The evaluation of potential suppliers was not documented.
Complaints
Complaint files are not adequately maintained.
Failure to submit, distinct marking
You did not submit a required product report prior to the introduction of the product into commerce.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.
Failure to submit
You did not submit a required supplemental report, prior to the introduction of the product into commerce, for a new or modified model within a model or chassis family.
Show 1 more citation
Failure to submit
You did not submit an annual report for products requiring one.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Jan 9, 2026. Each one asked for a Form 483.
Inspection ending May 2, 2025OAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 31, 2025. FDAgent holds no copy that FDA posted.
- Jan 9, 2026FDA-FOIA-2026-290Form 483Records released
- Jul 10, 2025FDA-FOIA-2025-8502Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files4 of 4
- 21 CFR 820.184Device history record2 of 4
- 21 CFR 1002.13Annual reports2 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.50Purchasing controls1 of 4
- 21 CFR 820.70Production and process controls1 of 4
- 21 CFR 820.100Corrective and preventive action1 of 4
- 21 CFR 1002.10Product reports1 of 4
Show all 12
- 21 CFR 1002.11Supplemental reports1 of 4
- 21 CFR 1010.3Identification1 of 4
- 21 CFR 1040.10Laser products1 of 4
- 21 CFR 1040.11Specific purpose laser products1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003067608 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003067608. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.