Skip to content

Focus Medical, LLC

23 Francis J Clarke Cir, Bethel, Connecticut, United States

FEI 3003067608

FDA has 6 inspections on file for Focus Medical, LLC in Bethel, Connecticut (FEI 3003067608), from 2010 to 2025. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 8 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.198(a), 21 CFR 820.184 and 21 CFR 1002.13.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
6, 2010–2025
1 OAI · 3 VAI · 2 NAI
Inspections with 483 observations shown
4 of 6
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Jan 9, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety Act8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation of device failures

      Complaints involving the possible failure of to meet any of its specifications were not investigated where necessary.

    2. Aperture label

      A label bearing the wording: "Laser aperture". is not affixed to the medical laser product, in close proximity to the aperture through which was emitted accessible laser radiation in excess of the accessible emission limits of Class I.

    3. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.

    4. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.

    5. Class IV "Danger" label

      Each Class IV laser product does not have affixed a label bearing the DANGER logotype specified in the regulation, bearing the wording "LASER RADIATION--AVOID EYE OR SKIN EXPOSURE TO DIRECT OR SCATTERED RADIATION" in the position specified in the regulation and bearing the wording "CLASS IV LASER PRODUCT" in the position specified in the regulation.

    Show 3 more citations
    1. ID label lacks name and address

      An identification label failed to provide the name and address of the.

    2. Legible reproductions

      Legible reproductions of all labels and hazard warnings required by the regulations were not provided with the laser product.

    3. Calibration of measurement system

      The Class IV medical laser product is not supplied with instructions specifying a procedure and schedule for calibration of the measurement system.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Failure to submit

      You did not submit an annual report for products requiring one.

    2. Personnel

      Personnel do not have the necessary training to perform their jobs.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documented evaluation

      The evaluation of potential suppliers was not documented.

    2. Complaints

      Complaint files are not adequately maintained.

    3. Failure to submit, distinct marking

      You did not submit a required product report prior to the introduction of the product into commerce.

    4. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.

    5. Failure to submit

      You did not submit a required supplemental report, prior to the introduction of the product into commerce, for a new or modified model within a model or chassis family.

    Show 1 more citation
    1. Failure to submit

      You did not submit an annual report for products requiring one.

  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Jan 9, 2026. Each one asked for a Form 483.

  • Inspection ending May 2, 2025OAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 31, 2025. FDAgent holds no copy that FDA posted.

    1. Jan 9, 2026FDA-FOIA-2026-290Form 483Records released
    2. Jul 10, 2025FDA-FOIA-2025-8502Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.

Show all 12

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003067608 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003067608. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.