Skip to content

Transdermal Cap Inc.

26 Alpha Park, Highland Heights, Ohio, United States

FEI 3008699663

FDA has 5 inspections on file for Transdermal Cap Inc. in Highland Heights, Ohio (FEI 3008699663), from 2013 to 2025. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection. They are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.

The most recent OAI inspection ended on Jun 30, 2015. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Seven regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
5, 2013–2025
1 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
5 of 5
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
3 requests, latest Jan 6, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ID label lacks place, month & year of manufacture

      An identification label fails to provide the month and year of manufacture.

    2. Failure to submit

      You did not submit an annual report for products requiring one.

    3. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

    4. Reproduction of affixed information

      The manufacturer did not provide or cause to be provided in all descriptive brochures pertaining to each laser product, a legible reproduction of the class designation and the warning required b to be affixed to the product.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality Audit/Reaudit - conducted

      Quality audits and reaudits have not been performed.

    2. Design history file

      The design history file was not established.

    3. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    4. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    5. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers were not evaluated and selected based on their ability to meet specified requirements.

    Show 6 more citations
    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    2. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated and maintained have not been adequately established.

    3. Supplier oversight

      The type and extent of control to be exercised over the product and contractors was not clearly defined.

    4. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.

    5. Acceptable supplier records, inadequate records

      Records of acceptable suppliers, contractors and consultants have not been adequately established.

    6. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    3. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

    4. Product rework adverse effects {see also 820.184}

      Documentation of rework and reevaluation activities does not include a determination of whether there has been any adverse effect from rework upon the product.

    5. Equipment control activity documentation

      Equipment calibrations, checks and maintenance activities have not been documented.

    Show 1 more citation
    1. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been adequately established.

    4. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    5. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    Show 2 more citations
    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    2. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality System Procedures

      Quality system procedures and instructions have not been established.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    4. Design history file

      The design history file was not established.

    5. Complaints

      Complaint files are not maintained.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Jan 6, 2026. Each one asked for a Form 483.

  • Inspection ending Jun 6, 2025VAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 8, 2025. FDAgent holds no copy that FDA posted.

    1. Jan 6, 2026FDA-FOIA-2026-131Form 483Records released
    2. Aug 11, 2025FDA-FOIA-2025-9606Form 483· 1 of 5 sitesRecords released
  • Inspection ending Mar 26, 2024NAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Aug 9, 2024FDA-FOIA-2024-7123Form 483· 1 of 5 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for this facility’s 5 inspections, with how many of them cited each.

Show all 11

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3008699663 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3008699663. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.