Med Pen Concepts, LLC
800 W Cummings Park Ste 5650, Woburn, Massachusetts, United States
FEI 3015602723
FDA has 3 inspections on file for Med Pen Concepts, LLC in Woburn, Massachusetts (FEI 3015602723), from 2022 to 2024. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 12 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Three of the four FOIA requests for this site have asked FDA for this inspection's 483.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.198(a).
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 3, 2022–2024
- 2 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Jul 31, 2025
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Complaints
Complaint files are not adequately maintained.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Show 7 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Lack of or inadequate In-process acceptance procedures
Procedures for the of in-process product have not been established.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
Supplier notification of changes
There is no agreement with suppliers to notify you of changes in the product or service.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Sampling plans
Sampling plans are not written and based on valid statistical rationale.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
DHR documentation of label release {see also 820.184}
The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.
Management review dates
The results and/or dates of management reviews are not documented.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Complaints
Complaint files are not adequately maintained.
Show 3 more citations
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Jul 31, 2025. Each one asked for a Form 483.
Inspection ending Jan 24, 2024OAI
3 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Jul 31, 2025FDA-FOIA-2025-9285Form 483Closed
- Jul 30, 2024FDA-FOIA-2024-6621Form 483· 1 of 4 sitesClosed
- Jun 17, 2024FDA-FOIA-2024-5222Form 483· 1 of 3 sitesClosed
Inspection ending Apr 13, 2022OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 4, 2023FDA-FOIA-2023-8608Form 483· 1 of 8 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
Show all 12
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.120Device labeling1 of 2
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 820.250Statistical techniques1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3015602723 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3015602723. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.