Skip to content

CC Wellness, LLC

24903 Avenue Kearny, Valencia, California, United States

FEI 3013395518DUNS 067692292

FDA has 5 inspections on file for CC Wellness, LLC in Valencia, California (FEI 3013395518), from 2014 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 6 citations from this inspection, all under 21 CFR Part 211, the manufacturing rules for finished drugs. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Four regulations were cited at more than one inspection; they are listed below.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, through Dec 31, 2027. Its registrant is CC Wellness, LLC (D-U-N-S 067692292).

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
5, 2014–2026
0 OAI · 4 VAI · 1 NAI
Inspections with 483 observations shown
4 of 5
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Oct 13, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Component identification test

      Specific identification tests are not conducted on components that have been accepted based on the supplier's report of analysis.

    2. Written in-process control procedures

      Written procedures are not established that describe the in-process controls, tests and examinations to be conducted on appropriate samples of in-process materials of each batch.

    3. Absence of Written Procedures

      Your firm failed to establish written procedures for production and process controls designed to assure that the drug products have the identity, strength, purity, and quality that they are purported or represented to possess.

    4. Component identity verification

      Drug product component testing is deficient in that at least one specific test to verify the identity of each component is not performed.

    5. Manufacturing Instructions and Specifications

      The master production and control records are deficient in that they do not include complete manufacturing and control instructions, sampling procedures, testing procedures, specifications, special notations and precautions to be followed.

    Show 1 more citation
    1. Written procedures not established/followed

      Written procedures are not followed for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Written record of investigation incomplete

      Written records of investigations into unexplained discrepancies and the failure of a batch or any of its components to meet specifications do not include the conclusions and follow-up.

    2. Absence of Written Procedures

      Your firm failed to establish adequate written procedures for production and process controls designed to assure that the drug products have the identity, strength, purity, and quality that they are purported or represented to possess.

    3. Written procedures not established/followed

      Written procedures are not established and followed for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product.

    4. Approve or reject procedures or specs

      The quality control unit lacks responsibility to approve and reject all procedures or specifications impacting on the identity, strength, quality and purity of drug products.

    5. Records reviewed annually

      Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures.

  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    2. Design history file

      The design history file was not established.

    3. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

    4. Document change records, maintained.

      Records of changes to documents were not adequately maintained.

    5. Investigation of device failures

      Complaints involving the possible failure of a device, labeling and packaging to meet any of its specifications were not reviewed, evaluated and investigated where necessary.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate DHR procedures

      Procedures for device history records have not been established.

    2. Results not used for expiration dates, storage cond.

      Results of stability testing are not used in determining expiration dates.

    3. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    4. DMR - not or inadequately maintained

      A device master record has not been maintained.

    5. Control procedures to monitor and validate performance

      Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.

    Show 2 more citations
    1. Cleaning / Sanitizing / Maintenance

      Equipment and utensils are not cleaned, maintained, and sanitized at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.

    2. Design control - no procedures

      Procedures for design control have not been established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Oct 13, 2026. Each one asked for a Form 483.

  • Inspection ending Aug 20, 2026VAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 27, 2026. FDAgent holds no copy that FDA posted.

    1. Oct 13, 2026FDA-FOIA-2026-9965Form 483· 1 of 5 sitesRecords released
  • Inspection ending Nov 15, 2023VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Feb 14, 2024FDA-FOIA-2024-1662Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 16 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.

Show all 16

FDA drug establishment registration

FDA registration search
Registered operations
Manufacturing
Registration runs through
Dec 31, 2027
Registrant
CC Wellness, LLCD-U-N-S 067692292

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3013395518 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3013395518. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.