Atrium Medical Corporation
40 Continental Blvd, Merrimack, New Hampshire, United States
FEI 3011175548
FDA has 9 inspections on file for Atrium Medical Corporation in Merrimack, New Hampshire (FEI 3011175548), from 2009 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 8 citations from this inspection: 6 under 21 CFR Part 820, the device quality system rules, and 2 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. Two of the eight FOIA requests for this site have asked FDA for this inspection's 483, and four more for the 483 from a later inspection, ending Jun 1, 2026, that FDA's Data Dashboard doesn't list yet.
The most recent OAI inspection ended on Feb 10, 2020. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 9, 2009–2024
- 4 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 4 of 9
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 8 requests, latest Sep 16, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
9 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device and labeling to meet any of its specifications were not reviewed and investigated where necessary.
Show 3 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Software validation for automated processes
Software used as part of production has not been validated for its intended use according to an established protocol.
Show 3 more citations
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Complaints
Complaint files are not adequately maintained.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Info evaluated to determine if event was reportable
The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.
Investigation of device failures
Complaints involving the possible failure of a device and packaging to meet any of its specifications were not investigated where necessary.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Incoming inspection, testing, verification
Incoming product was not adequately inspected or tested to verify conformance to specifications.
Documentation
Process validation activities and results have not been fully documented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Eight requests to FDA under the Freedom of Information Act named this site, most recently on Sep 16, 2026. Each one asked for a Form 483.
Inspection ending Jun 1, 2026
4 requests. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Jul 20, 2026. FDAgent holds no copy that FDA posted.
- Sep 16, 2026FDA-FOIA-2026-8970Form 483Records released
- Aug 21, 2026FDA-FOIA-2026-8138Form 483· 1 of 4 sitesOpen
- Aug 12, 2026FDA-FOIA-2026-7774Form 483· 1 of 4 sitesRecords released
- Jun 30, 2026FDA-FOIA-2026-6214Form 483Records released
Inspection ending Nov 26, 2024VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 11, 2025. FDAgent holds no copy that FDA posted.
- Oct 15, 2025FDA-FOIA-2025-12015Form 483· 1 of 5 sitesOpen
- May 15, 2025FDA-FOIA-2025-4437Form 483· 1 of 5 sitesRecords released
Inspection date not given
2 requests.
- Jul 23, 2026FDA-FOIA-2026-7109Form 483Open
- Jan 17, 2025FDA-FOIA-2025-502Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 9 inspections, with how many of them cited each.
- 21 CFR 820.75Process validation3 of 4
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 4
- 21 CFR 820.70Production and process controls2 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 4
- 21 CFR 820.90Nonconforming product2 of 4
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 4
Show all 9
- 21 CFR 820.30Design controls1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3011175548 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3011175548. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.