Tactical Medical Solutions, Thermal Management Group
582 S Econ Cir, Oviedo, Florida, United States
FEI 3009141585
FDA has 2 inspections on file for Tactical Medical Solutions, Thermal Management Group in Oviedo, Florida (FEI 3009141585), from 2025 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2025–2026
- 0 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Feb 10, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.4.1
Supplier performance not monitored
Supplier performance in meeting requirements for the purchased product was not monitored.
- ISO 13485:2016 Clause 8.5.2
Corrective action not taken
Corrective action was not adequately taken to eliminate the cause of nonconformities in order to prevent recurrence.
- ISO 13485:2016 Clause 5.6.1
QMS not reviewed by top management at documented planned intervals
Top management did not adequately review the quality management system at documented planned intervals to ensure its continued suitability, adequacy, and effectiveness.
- ISO 13485:2016 Clause 8.3.1
Records of nonconforming product and actions taken not maintained
Records of the nature of nonconforming product and any subsequent action taken, including the evaluation, and any investigation and the rationale for decisions, were not adequately maintained.
- ISO 13485:2016 Clause 6.2(b)
Personnel training not provided
Training has not been provided, or other actions needed taken, to achieve or maintain the necessary personnel competence.
Show 1 more citation
- ISO 13485:2016 Clause 8.2.2
Complaint handling records not maintained
Complaint handling records were not adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices14 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Lack of or inadequate In-process acceptance procedures
Procedures for the acceptance of in-process product have not been adequately established.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
Show 9 more citations
Quality System Procedures
Quality system procedures and instructions have not been established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Design control - no procedures
Procedures for design control have not been established.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis was not performed.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Feb 10, 2026. Each one asked for a Form 483.
Inspection ending Aug 8, 2025VAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Nov 10, 2025. FDAgent holds no copy that FDA posted.
- Feb 10, 2026FDA-FOIA-2026-1423Form 483· 1 of 3 sitesRecords released
- Sep 29, 2025FDA-FOIA-2025-11357Form 483Records released
- Sep 18, 2025FDA-FOIA-2025-11060Form 483· 1 of 4 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 801.20Label to bear a unique device identifier1 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.25Personnel1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
Show all 12
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.198Complaint files1 of 2
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3009141585 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3009141585. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.