ROMD, LLC
2967 Waterview Dr, Rochester Hills, Michigan, United States
FEI 3008959459
FDA has 5 inspections on file for ROMD, LLC in Rochester Hills, Michigan (FEI 3008959459), from 2012 to 2024. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.75(a) and 21 CFR 820.30(i).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2012–2024
- 0 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 5 of 5
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Design output - documentation
Design output was not documented before release.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Process changes - review, evaluation and revalidation
A validated process was not reviewed and evaluated and revalidated when changes or process deviations occurred.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Process validation activities and results have not been documented and approved.
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
DHR documentation of label release {see also 820.184}
The DHR does not include complete records of examination and release of device labeling, including date and signature of the examiner.
Equipment control activity documentation
Equipment maintenance activities have not been documented.
Show 4 more citations
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Approval, inadequate purchasing data
Purchasing data that clearly describe or reference specified requirements for purchased or otherwise received product and services have not been adequately approved and adequately established.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls3 of 5
- 21 CFR 820.75Process validation3 of 5
- 21 CFR 820.50Purchasing controls2 of 5
- 21 CFR 820.25Personnel1 of 5
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 5
- 21 CFR 820.120Device labeling1 of 5
- 21 CFR 820.181Device master record1 of 5
Show all 9
- 21 CFR 820.198Complaint files1 of 5
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3008959459 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3008959459. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.