Skip to content

The Standing Company

5848 Dixie Hwy, Saginaw, Michigan, United States

FEI 3007494904

FDA has 5 inspections on file for The Standing Company in Saginaw, Michigan (FEI 3007494904), from 2010 to 2015. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Mar 27, 2013. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.198(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
5, 2010–2015
2 OAI · 2 VAI · 1 NAI
Inspections with 483 observations shown
3 of 5
Form 483s posted by FDA
None held for this FEI
Warning letters
· CDRH ↗
2 on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Results of investigation

      Records of complaint investigations do not include the dates and results of the investigation.

    2. Design control - no procedures

      Procedures to control the design process of the device were not complete and implemented.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Periodic equipment inspection lack of or inadequate procedu

      Periodic inspections of equipment were not conducted in accordance with established procedures to ensure adherence to applicable maintenance schedules were not documented.

    2. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    3. Design history file

      The design history file was not established or maintained and does not demonstrate that the design was developed following the approved design plan and the design control requirements of 21 CFR 820.

    4. Management review - defined interval, sufficient frequency

      Management reviews were not conducted at defined intervals.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures were not established for the identification, documentation, validation or verification, review, and approval of design changes before their implementation.

    Show 6 more citations
    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not implemented.

    2. Lack of procedures, or not maintained

      Document control procedures were not complete.

    3. Implement quality policy

      Not all employees know there is and understand the quality policy.

    4. General

      Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been implemented.

    5. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not complete.

    6. Statistical methodology

      Corrective and preventive action procedures addressing the use of appropriate statistical methodology to identify existing and potential causes of nonconforming product or other quality problems were not established.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.

Show all 10

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3007494904 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3007494904. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.