Dexcowin Co. Ltd.
901~905 25 Gasan Digital 1-Ro, Geumcheon, Korea (the Republic of)
FEI 3007065358
FDA has 5 inspections on file for Dexcowin Co. Ltd. in Geumcheon, Korea (the Republic of) (FEI 3007065358), from 2011 to 2023. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA posted the Form 483 from that inspection. FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below. Two FOIA requests have asked FDA for this inspection's 483.
The most recent OAI inspection ended on Mar 26, 2015. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). 2 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2011–2023
- 1 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 5
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 2 requests, latest Feb 11, 2026
- Import refusals
- 2 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesPosted 483 · 1 citation
Read this Form 483 on its own page →Form 483 PDF on fda.gov
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Software validation for automated processes
Software used as part of production has not been adequately validated for its intended use according to an established protocol.
Certification not based on adequate test/testing program
Certification was not based upon a testing program in accordance with good manufacturing practices.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established and maintained.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 11 more citations
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.
Failure to report
You did not immediately report to the Director, CDRH, FDA, an accidental radiation occurrence reported to or otherwise known to you. involving a product introduced or intended to be introduced into commerce by you.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been maintained and implemented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Maintenance schedule (all x-ray eqpt)
Manuals or instruction sheets which include a schedule of the maintenance necessary to keep the equipment in compliance with the regulations were not provided to purchasers.
Certification not based on adequate test/testing program
Certification was not based upon a testing program in accordance with good manufacturing practices.
Training records
Personnel training is not documented.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Show 3 more citations
Failure to submit
You did not submit an annual report for products requiring one.
Design input - documentation
Design input requirements were not adequately documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety ActCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Feb 11, 2026. Each one asked for a Form 483.
Inspection ending Feb 23, 2023VAI
2 requests. In the inspection history above. FDA has posted its Form 483.
- Feb 11, 2026FDA-FOIA-2026-1454Form 483· 1 of 2 sitesRecords released
- Dec 16, 2025FDA-FOIA-2025-14106Form 483· 1 of 2 sitesOpen
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 17 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files3 of 3
- 21 CFR 820.25Personnel2 of 3
- 21 CFR 820.30Design controls2 of 3
- 21 CFR 820.184Device history record2 of 3
- 21 CFR 1010.2Certification2 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.22Quality audit1 of 3
Show all 17
- 21 CFR 820.40[Reserved]1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.75Process validation1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
- 21 CFR 820.200Servicing1 of 3
- 21 CFR 1002.13Annual reports1 of 3
- 21 CFR 1002.20Reporting of accidental radiation occurrences1 of 3
- 21 CFR 1020.30Diagnostic x-ray systems and their major components1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3007065358, one row per refused product line.
2 refusals from 2018-07-13 to 2018-09-30 · charged FAILS STD on 1, REGISTERED on 1 · no FDA sample analysis
FDA firm profile- 2018-09-30
GENERAL PURPOSE LASER PRODUCTS, NON-MEDICAL95RDW
FAILS STD 501(e), 801(a)(3); ADULTERATION
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-6262932-2/11/1
- 2018-07-13
SYSTEM, X-RAY, EXTRAORAL SOURCE, DIGITAL90MUH
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-6578788-7/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3007065358 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3007065358. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.