Skip to content

Zyno Medical LLC

177 Pine St, Natick, Massachusetts, United States

FEI 3006575795

FDA has 7 inspections on file for Zyno Medical LLC in Natick, Massachusetts (FEI 3006575795), from 2008 to 2025. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published one citation from this inspection, under 21 CFR Part 803, the device adverse event reporting rules. It is listed below. One of the four FOIA requests for this site has asked FDA for this inspection's 483.

The most recent OAI inspection ended on Feb 15, 2022. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.

Six regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2008–2025
2 OAI · 4 VAI · 1 NAI
Inspections with 483 observations shown
6 of 7
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
4 requests, latest Oct 31, 2025
Warning letters
· CDRH ↗
2 on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    2. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.

    3. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documented evaluation

      The evaluation of potential suppliers was not documented.

    2. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    3. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    4. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    Show 1 more citation
    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - software validation documentation

      Results of the validation of the device software were not adequately documented.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    4. Servicing - Lack of or inadequate procedures

      Procedures or instructions for performing servicing activities have not been adequately established.

    5. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    Show 3 more citations
    1. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    2. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

  7. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Four requests to FDA under the Freedom of Information Act named this site, most recently on Oct 31, 2025. Each one asked for a Form 483.

  • Inspection ending Sep 18, 2025VAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Mar 16, 2026. FDAgent holds no copy that FDA posted.

    1. Oct 31, 2025FDA-FOIA-2025-12503Form 483· 1 of 5 sitesRecords released
  • Inspection ending Aug 16, 2023VAI

    2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Mar 29, 2024FDA-FOIA-2024-2958Form 483Closed
    2. Nov 2, 2023FDA-FOIA-2023-9650Form 483Closed
  • Inspection ending Feb 15, 2022OAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Mar 29, 2024FDA-FOIA-2024-2957Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 7 inspections, with how many of them cited each.

Show all 11

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3006575795 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3006575795. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.