Skip to content

Generic Medical Devices, Inc.

5727 Baker Way NW Ste 201, Gig Harbor, Washington, United States

FEI 3006142121

FDA has 2 inspections on file for Generic Medical Devices, Inc. in Gig Harbor, Washington (FEI 3006142121), from 2010 to 2012. The latest ended on and was classified OAI, official action indicated. FDA has issued 2 warning letters under this FEI, the latest on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 11 citations from this inspection: 8 under 21 CFR Part 820, the device quality system rules, and 3 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

No court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 803.50(a)(1) and 21 CFR 820.100(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2010–2012
2 OAI · 0 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Warning letters
· CDRH ↗
2 on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Process changes - review, evaluation and revalidation

      A validated process was not revalidated when changes or process deviations occurred.

    2. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Supplier oversight

      The type and extent of control to be exercised over the product, services, suppliers, and contractors was not clearly defined.

    5. Approval, inadequate purchasing data

      Purchasing data that clearly describe or reference specified requirements for purchased or otherwise received product and services have not been adequately established.

    Show 6 more citations
    1. Document change records, maintained.

      Records of changes to documents were not adequately maintained.

    2. Submission of Supplement to Individual Report Within One Month

      A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.

    3. Documented evaluation

      The evaluation of potential contractors and consultants was not documented.

    4. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    5. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    6. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    4. Evaluation, timeliness, identification

      Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.

    5. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    Show 1 more citation
    1. Documentation

      Corrective and preventive action activities and/or results have not been documented.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

  • 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 2
  • 21 CFR 820.100Corrective and preventive action2 of 2
  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
  • 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 2
  • 21 CFR 820.20Management responsibility1 of 2
  • 21 CFR 820.30Design controls1 of 2
  • 21 CFR 820.40[Reserved]1 of 2
  • 21 CFR 820.50Purchasing controls1 of 2
Show all 11

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3006142121 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3006142121. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.