audifon GmbH & Co. KG
Werner-Von-Siemens-Str. 2, Kolleda, Germany
FEI 3006076696
FDA has 3 inspections on file for audifon GmbH & Co. KG in Kolleda, Germany (FEI 3006076696), from 2017 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2017–2025
- 0 OAI · 3 VAI · 0 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Oct 14, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Show 1 more citation
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Design review - Lack of or inadequate procedures
Procedures for design review have not been established.
Show 4 more citations
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Oct 14, 2025. It asked for a Form 483.
Inspection ending Jan 16, 2025VAI
1 request. In the inspection history above.
- Oct 14, 2025FDA-FOIA-2025-11830Form 483Open
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 820.30Design controls2 of 3
- 21 CFR 820.50Purchasing controls2 of 3
- 21 CFR 820.75Process validation2 of 3
- 21 CFR 820.184Device history record2 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 3
Show all 10
- 21 CFR 820.198Complaint files1 of 3
- 21 CFR 820.200Servicing1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006076696 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006076696. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.