TELEFLEX LLC
3015 Carrington Mill Blvd, Morrisville, North Carolina, United States
FEI 3005747797
FDA has 8 inspections on file for TELEFLEX LLC in Morrisville, North Carolina (FEI 3005747797), from 2010 to 2022. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
4 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.100(b), 21 CFR 820.90(a) and 21 CFR 820.90(b)(2).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 8, 2010–2022
- 0 OAI · 4 VAI · 4 NAI
- Inspections with 483 observations shown
- 4 of 8
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 4 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Production processes
Production processes were not monitored to ensure that a device conforms to its specifications.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Supplier oversight
The type and extent of control to be exercised over the product, services, and contractors was not clearly defined.
Complaints
Complaint files are not adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately and fully validated and approved according to established procedures.
Product rework documentation, DHR {see also 820.184}
Rework and reevaluation activities have not been fully documented in the device history record.
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Investigation
Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.
Show 2 more citations
DHR content
The device history record does not demonstrate the device is manufactured in accordance with the device master record.
Nonconforming product, Lack of or inadequate procedures
Procedures were not complete for the control of products that do not conform to specifications.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action4 of 4
- 21 CFR 820.90Nonconforming product3 of 4
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 4
- 21 CFR 820.30Design controls1 of 4
- 21 CFR 820.50Purchasing controls1 of 4
- 21 CFR 820.70Production and process controls1 of 4
- 21 CFR 820.75Process validation1 of 4
- 21 CFR 820.184Device history record1 of 4
Show all 9
- 21 CFR 820.198Complaint files1 of 4
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3005747797, one row per refused product line.
4 refusals from 2019-04-29 to 2022-04-20 · charged NOT LISTED on 4, FRNMFGREG on 4 · no FDA sample analysis
FDA firm profile- 2022-04-20
SAW, MANUAL AND ACCESSORIES79GDR
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsSCS-0861554-7/11/1
- 2021-04-14
CATHETER, PERCUTANEOUS, INTRAVASCULAR, LONG TERM80LJS
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsSCS-4339803-3/21/1
- 2021-04-14
CATHETER, PERCUTANEOUS, INTRAVASCULAR, LONG TERM80LJS
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsSCS-4339803-3/11/1
- 2019-04-29
SET, ADMINISTRATION, INTRAVASCULAR80FPA
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsSCS-4584191-4/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005747797 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005747797. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.